INSTITUTE FOR TESTING AND CERTIFICATION

Scope of accreditation ACL 2222

RELATED LINKS


The current scope of accreditation of Accredited Calibration Laboratory No. 2222


(note: content is not optimized for mobile devices, we recommend switching to "desktop" view)

The Appendix is an integral part of Certificate of Accreditation No. 21/2022 of 07/01/2022
Accredited entity according to ČSN EN ISO/IEC 17025:2018
Institut pro testování a certifikaci, a. s., Calibration Laboratory, Sokolovská 573, 686 01 Uherské Hradiště


Calibration laboratory locations:

1.       Uherské Hradiště                             Sokolovská 573, 686 01 Uherské Hradiště

2.       Zlín                                                   třída Tomáše Bati 5264, 760 01 Zlín

3.       Praha                                               Křelovická 970, 104 00 Praha 10 – Uhříněves

4.       Brno                                                Křižíkova 70, 612 00 Brno

 

CMC for the field of measured quantity:  Length

Ord. number1

Calibrated quantity / Subject of calibration

Nominal range

Parameter(s) of the meas. quantity

Lowest expanded measurement uncertainty specified2

Calibration principle

Calibration procedure identification3

Work-place

min.

unit

 

max.

unit

1

Slide gauges

0

mm

to

100

mm

 

 

 

12 mm

Direct measurement using parallel gauge blocks

KP 1/D

1

 

 

100

mm

to

500

mm

 

 

 

14 mm

 

 

 

2

Depth gauges

0

mm

to

200

mm

 

 

 

13 mm

Direct measurement using parallel gauge blocks

KP 1/D

 

3

Height gauges

0

mm

to

500

mm

 

 

 

14 mm

Direct measurement using parallel gauge blocks

KP 1/D

 

4

Micrometers

0

mm

to

50

mm

 

 

 

1.6 mm

Direct measurement using parallel gauge blocks

KP 2/D

 

 

 

50

mm

to

100

mm

 

 

 

2.0 mm

 

 

 

 

 

100

mm

to

200

mm

 

 

 

3.2 mm

 

 

 

5

Micrometer depth gauges

0

mm

to

100

mm

 

 

 

2 mm

Direct measurement using parallel gauge blocks

KP 2/D

 

6

Three-contact internal gauge with a micrometer screw

0

mm

to

50

mm

 

 

 

1.6 mm

Direct measurement of standard measuring rings.

KP 2/D

 

 

 

50

mm

to

75

mm

 

 

 

1.8 mm

 

 

 

 

 

75

mm

to

100

mm

 

 

 

2.0 mm

 

 

 

7

Three-contact internal gauge with a lever mechanism

0

mm

to

50

mm

 

 

 

2.1 mm

Direct measurement of standard measuring rings

KP 2/D

 

 

 

50

mm

to

75

mm

 

 

 

2.3 mm

 

 

 

 

 

75

mm

to

100

mm

 

 

 

2.5 mm

 

 

 

8

Two-contact internal gauge

0

mm

to

35

mm

 

 

 

2.5 mm

Direct measurement of standard measuring rings

KP 2/D

 

 

 

35

mm

to

65

mm

 

 

 

2.6 mm

 

 

 

 

 

65

mm

to

75

mm

 

 

 

2.7 mm

 

 

 

 

 

75

mm

to

100

mm

 

 

 

2.9 mm

 

 

 

9

Pasameters

0

mm

to

25

mm

 

 

 

0.5 mm

Direct measurement using parallel gauge blocks

KP 2/D

1

 

 

25

mm

to

50

mm

 

 

 

0.8 mm

 

 

 

 

 

50

mm

to

75

mm

 

 

 

1.2 mm

 

 

 

 

 

75

mm

to

100

mm

 

 

 

1.5 mm

 

 

 

 

 

100

mm

to

125

mm

 

 

 

1.8 mm

 

 

 

 

 

125

mm

to

150

mm

 

 

 

2.2 mm

 

 

 

 

 

150

mm

to

175

mm

 

 

 

2.5 mm

 

 

 

 

 

175

mm

to

200

mm

 

 

 

2.9 mm

 

 

 

10

Micropasameters

0

mm

to

25

mm

 

 

 

2.0 mm

Direct measurement using parallel gauge blocks

KP 2/D

 

 

 

25

mm

to

50

mm

 

 

 

2.5 mm

 

 

 

 

 

50

mm

to

75

mm

 

 

 

3.2 mm

 

 

 

 

 

75

mm

to

100

mm

 

 

 

3.8 mm

 

 

 

 

 

100

mm

to

125

mm

 

 

 

4.5 mm

 

 

 

 

 

125

mm

to

150

mm

 

 

 

5.4 mm

 

 

 

 

 

150

mm

to

175

mm

 

 

 

6.1 mm

 

 

 

 

 

175

mm

to

200

mm

 

 

 

6.9 mm

 

 

 

11

Centesimal indicators

0

mm

to

100

mm

 

 

(2.L

+6) mm

Comparison with a standard length gauge

KP 3/D

 

12

Millesimal indicators

0

mm

to

50

mm

 

 

(2.L

+1) mm

Comparison with a standard length gauge

KP 3/D

 

13

Cylindrical gauges

0

mm

to

200

mm

 

 

(3.L

+0.5) mm

Direct measurement by a length gauge

KP 3/D

 

14

Slot gauges

0

mm

to

200

mm

 

 

(3.L

+0.5) mm

Direct measurement by a length gauge

KP 3/D

 

15

Limit plug gauges

0

mm

to

200

mm

 

 

(5.L

+3) mm

Direct measurement by a length gauge

KP 3/D

 

16

Precision plug gauges

0

mm

to

50

mm

 

 

(1.L

+0.5) mm

Direct measurement by a length gauge

KP 3/D

 

17

Measuring wires

0.17

mm

to

6.35

mm

 

 

(1.L

+0.5) mm

Direct measurement by a length gauge

KP 3/D

 

18

Setting rings

10

mm

to

200

mm

 

 

(4.L

+1.2) mm

Direct measurement by a length gauge

KP 3/D

 

19

Micrometer depth gauges

0

mm

to

100

mm

 

 

 

2.0 mm

Direct measurement by a length gauge

KP 3/D

 

20

Steel rules

0

mm

to

250

mm

 

 

 

0.05 mm

Direct measurement by a standard microscope

KP 4/D

 

21

Rules – straightness

0

mm

to

250

mm

 

 

 

0.007 mm

Direct measurement by a standard microscope.

KP 4/D

 

22

Thread gauges

0

mm

to

300

mm

 

 

 

0.01 mm

Direct measurement by a standard microscope.

KP 4/D

1

23

Radius gauges

0

mm

to

300

mm

 

 

 

0.01 mm

Direct measurement by a standard microscope.

KP 4/D

 

24

Profile gauges

0

mm

to

300

mm

 

 

(1.L

+0.5) mm

Direct measurement by a standard microscope.

KP 4/D

 

25

Parallel gauge blocks

0.5

mm

to

100

mm

 

 

(2.L

+0.2) mm

Comparison with standard parallel gauge blocks

KP 21/D

 

                               

1)    Asterisk at the ordinal number identifies the calibrations, which the Laboratory is qualified to carry out outside the permanent laboratory premises.

2     The expanded measurement uncertainty is in accordance with ILAC-P14 and EA-4/02, part of CMC, and it is the lowest value of the respective uncertainty. If not stated otherwise, its coverage probability is approx. 95 %. If not stated otherwise, the uncertainty values stated without a unit are relative to the value measured. If the calibration is carried out outside the laboratory premises, the measurement uncertainty may be affected.

3     If the document identifying the calibration procedure is dated, only these specific procedures are used. If the document identifying the calibration procedure is not dated, the latest edition of the specified procedure is used (including any changes).

L    measured length in metres

 

CMC for the field of measured quantity:  Plane angle

Ord. number1

Calibrated quantity / Subject of calibration

Nominal range

Parameter(s) of the meas. quantity

Lowest expanded measurement uncertainty specified2

Calibration principle

Calibration procedure identification3

Work-place

min.

unit

 

max.

unit

1

Angles

0

°

to

180

°

 

 

 

2.5 ¢

Direct measurement by a standard microscope.

KP 4/D

1

2

Angle gauges

0

°

to

360

°

 

 

 

2.5 ¢

Direct measurement by a standard microscope.

KP 4/D

 

                             

1)    Asterisk at the ordinal number identifies the calibrations, which the Laboratory is qualified to carry out outside the permanent laboratory premises.

2     The expanded measurement uncertainty is in accordance with ILAC-P14 and EA-4/02, part of CMC, and it is the lowest value of the respective uncertainty. If not stated otherwise, its coverage probability is approx. 95 %. If not stated otherwise, the uncertainty values stated without a unit are relative to the value measured. If the calibration is carried out outside the laboratory premises, the measurement uncertainty may be affected.

3     If the document identifying the calibration procedure is dated, only these specific procedures are used. If the document identifying the calibration procedure is not dated, the latest edition of the specified procedure is used (including any changes).

 

CMC for the field of measured quantity:  Mechanical motion (speed)

Ord. number1

Calibrated quantity / Subject of calibration

Nominal range

Parameter(s) of the meas. quantity

Lowest expanded measurement uncertainty specified2

Calibration principle

Calibration procedure identification3

Work-place

min.

unit

 

max.

unit

1.

Speed / analogue and digital contact and contactless, revolution counters, rpm sensors, stroboscopes.

1

min-1

to

10,000

min-1

 

 

 

0.006 min-1

Direct measurement of optical signal or direct comparison with a standard revolution counter

KP 6/F

1

 

 

10,000

min-1

to

999,999

min-1

 

 

 

0.03 min-1

 

 

 

 

 

2

min-1

to

999.99

min-1

 

 

 

0.1 min-1

Measurement by a standard revolution counter

 

 

 

 

1,000

min-1

to

9,999

min-1

 

 

 

0.9 min-1

 

 

 

 

 

10,000

min-1

to

999,999

min-1

 

 

 

7.0 min-1

 

 

 

1)    Asterisk at the ordinal number identifies the calibrations, which the Laboratory is qualified to carry out outside the permanent laboratory premises.

2     The expanded measurement uncertainty is in accordance with ILAC-P14 and EA-4/02, part of CMC, and it is the lowest value of the respective uncertainty. If not stated otherwise, its coverage probability is approx. 95 %. If not stated otherwise, the uncertainty values stated without a unit are relative to the value measured. If the calibration is carried out outside the laboratory premises, the measurement uncertainty may be affected.

3     If the document identifying the calibration procedure is dated, only these specific procedures are used. If the document identifying the calibration procedure is not dated, the latest edition of the specified procedure is used (including any changes).

 

 

CMC for the field of measured quantity:  Hardness

Ord. number1

Calibrated quantity / Subject of calibration

Nominal range

Parameter(s) of the meas. quantity

Lowest expanded measurement uncertainty specified2

Calibration principle

Calibration procedure identification3

Work-place

min.

unit

 

max.

unit

1

Hardness tester Shore A

1

Shore A

to

100

Shore A

Spring force 31 mN

 

 

0.3 Shore A

Direct measurement of force on a calibrator and measurement of tip geometry using an image analyzer

KP 1/H

2

2

Hardness tester Shore C

1

Shore C

to

100

Shore C

Spring force 71 mN

 

 

0.3 Shore C

 

 

 

3

Hardness tester Shore D

1

Shore D

to

100

Shore D

Spring force 71 mN

 

 

0.3 Shore D

 

 

 

4

Hardness tester IRHD/N

10

IRHD

to

100

IRHD

 

 

 

0.6 IRHD

Direct measurement of force on a calibrator and measurement of tip geometry using an image analyzer

KP 2/H

 

5

Hardness tester IRHD/m

10

IRHD

to

100

IRHD

 

 

 

1.1 IRHD

 

 

 

                               

1     Asterisk at the ordinal number identifies the calibrations, which the Laboratory is qualified to carry out outside the permanent laboratory premises.

2     The expanded measurement uncertainty is in accordance with ILAC-P14 and EA-4/02, part of CMC, and it is the lowest value of the respective uncertainty. If not stated otherwise, its coverage probability is approx. 95 %. If not stated otherwise, the uncertainty values stated without a unit are relative to the value measured. If the calibration is carried out outside the laboratory premises, the measurement uncertainty may be affected.

3     If the document identifying the calibration procedure is dated, only these specific procedures are used. If the document identifying the calibration procedure is not dated, the latest edition of the specified procedure is used (including any changes).

 

 

CMC for the field of measured quantity:  Pressure, mechanical stress

Ord. number1

Calibrated quantity / Subject of calibration

Nominal range

Parameter(s) of the meas. quantity

Lowest expanded measurement uncertainty specified2

Calibration principle

Calibration procedure identification3

Work-place

min.

unit

 

max.

unit

1.

Deformation pressure gauges

0

kPa

to

10

kPa

Overpressure

gas

0.015 %

+ 0.058 kPa

Comparison with a standard digital  manometer

KP 1/P

1, 4

 

 

10

kPa

to

400

kPa

 

 

0.046 %

+ 0.07 kPa

 

 

 

 

 

0.4

MPa

to

2

MPa

 

 

0.046 %

+ 0.18 kPa

 

 

 

 

 

2

MPa

to

6

MPa

 

 

0.015 %

+ 0.42 kPa

 

 

 

 

 

0

kPa

to

95

kPa

Underpressure

gas

0.046 %

+ 0.07 kPa

 

 

 

 

 

70

kPa

to

130

kPa

Absolute pressure

gas4

0.015 %

+ 0.12 kPa

 

 

 

 

 

130

kPa

to

520

kPa

 

 

0.046 %

+ 0.13 kPa

 

 

 

 

 

0.52

MPa

to

2.12

MPa

 

 

0.046 %

+ 0.21 kPa

 

 

 

 

 

2.12

MPa

to

6

MPa

 

 

0.015 %

+ 0.44 kPa

 

 

 

 

 

6

MPa

to

16

MPa

Overpressure, absolute pressure

gas, liquid4

0.015 %

+ 1.2 kPa

 

 

 

 

 

16

MPa

to

60

MPa

Overpressure, absolute pressure

liquid4

0.046 %

+ 11 kPa

 

 

 

 

 

60

MPa

to

100

MPa

 

 

0.012 %

+ 8.3 kPa

 

 

 

 

 

70

kPa

to

120

kPa

Barometric pressure

 

 

0.082 kPa

 

 

 

2*

Pressure converters

0

kPa

to

10

kPa

Overpressure

gas

0.015 %

+ 0.0018 kPa

Comparison with a standard digital  manometer

KP 2/P

 

 

 

10

kPa

to

400

kPa

 

 

0.048 %

+ 0.035 kPa

 

 

 

 

 

0.4

MPa

to

2

MPa

 

 

0.048 %

+ 0.12 kPa

 

 

 

 

 

2

MPa

to

6

MPa

 

 

0.015 %

+ 0.35 kPa

 

 

 

 

 

0

kPa

to

95

kPa

Underpressure

gas

0.048 %

+ 0.035 kPa

 

 

 

 

 

70

kPa

to

130

kPa

Absolute pressure

gas4

0.015 %

+ 0.1 kPa

 

 

 

 

 

130

kPa

to

520

kPa

 

 

0.048 %

+ 0.11 kPa

 

 

 

 

 

0.52

MPa

to

2.12

MPa

 

 

0.048 %

+ 0.16 kPa

 

 

 

 

 

2.12

MPa

to

6

MPa

 

 

0.015 %

+ 0.36 kPa

 

 

 

 

 

6

MPa

to

16

MPa

Overpressure, absolute pressure

gas, liquid4

0.015 %

+ 0.93 kPa

 

 

 

 

 

16

MPa

to

60

MPa

Overpressure, absolute pressure

liquid4

0.048 %

+ 11 kPa

 

 

 

 

 

60

MPa

to

100

MPa

 

 

0.012 %

+ 8.1 kPa

 

 

 

 

 

70

kPa

to

120

kPa

Barometric pressure

 

0.12 %

 

 

 

 

3*

Digital manometers

0

kPa

to

10

kPa

Overpressure

gas

0.015 %

+ 0.0018 kPa

Comparison with a standard digital  manometer

KP 3/P

 

 

 

10

kPa

to

400

kPa

 

 

0.046 %

+ 0.039 kPa

 

 

 

 

 

0.4

MPa

to

2

MPa

 

 

0.046 %

+ 0.14 kPa

 

 

 

 

 

2

MPa

to

6

MPa

 

 

0.015 %

+ 0.36 kPa

 

 

 

 

 

0

kPa

to

95

kPa

Underpressure

gas

0.046 %

+ 0.039 kPa

 

 

 

 

 

70

kPa

to

130

kPa

Absolute pressure

gas4

0.015 %

+ 0.1 kPa

 

 

 

 

 

130

kPa

to

520

kPa

 

 

0.046 %

+ 0.11 kPa

 

 

 

 

 

0.52

MPa

to

2.12

MPa

 

 

0.046 %

+ 0.17 kPa

 

 

 

 

 

2.12

MPa

to

6

MPa

 

 

0.015 %

+ 0.37 kPa

 

 

 

 

 

6

MPa

to

16

MPa

Overpressure, absolute pressure

gas, liquid4

0.015 %

+ 0.98 kPa

 

 

 

 

 

16

MPa

to

60

MPa

Overpressure, absolute pressure

liquid4

0.046 %

+ 0.11 kPa

 

 

 

 

 

60

MPa

to

100

MPa

 

 

0.012 %

+ 8.2 kPa

 

 

 

 

 

70

kPa

to

120

kPa

Barometric pressure

 

 

0.058 kPa

 

 

 

1)    Asterisk at the ordinal number identifies the calibrations, which the Laboratory is qualified to carry out outside the permanent laboratory premises.

2     The expanded measurement uncertainty is in accordance with ILAC-P14 and EA-4/02, part of CMC, and it is the lowest value of the respective uncertainty. If not stated otherwise, its coverage probability is approx. 95 %. If not stated otherwise, the uncertainty values stated without a unit are relative to the value measured. If the calibration is carried out outside the laboratory premises, the measurement uncertainty may be affected.

3     If the document identifying the calibration procedure is dated, only these specific procedures are used. If the document identifying the calibration procedure is not dated, the latest edition of the specified procedure is used (including any changes).

4        The resulting pressure is the sum of relative and barometric pressure. The resulting uncertainty is the combination of relative and barometric pressure uncertainty.

 

CMC for the field of measured quantity:  Temperature

Ord. num-ber1

Calibrated quantity / Subject of calibration

Nominal range

Parameter(s) of the meas. quantity

Lowest expanded measurement uncertainty specified2

Calibration principle

Calibration procedure identification3

Work-place

min.

unit

 

max.

unit

1*

Thermocouples

-40

°C

to

0

°C

 

 

 

0.6 °C

Comparison with a standard resistance sensor in liquid bath.

KP 1/T

1, 4

 

 

 

 

 

0

°C

 

 

 

0.6 °C

Comparison with a standard resistance sensor in dry block.

 

 

 

 

0

°C

to

100

°C

 

 

 

0.6 °C

 

 

 

 

 

100

°C

to

250

°C

 

 

 

0.7 °C

 

 

 

 

 

250

°C

to

600

°C

 

 

 

0.8 °C

 

 

 

 

 

600

°C

to

800

°C

 

 

 

1.5 °C

Comparison with a standard thermoelectric sensor in a horizontal oven

 

 

 

 

800

°C

to

1,200

°C

 

 

 

2.2 °C

 

 

 

2*

Resistance temperature sensors, resistance temperature sensors with a transducer

-80

°C

to

-40

°C

 

 

 

0.20 °C

Comparison with a standard thermoelectric sensor in dry block

KP 2/T

4

 

 

-40

°C

to

0

°C

 

 

 

0.07 °C

Comparison with a standard resistance sensor in liquid bath.

 

 

 

 

 

 

 

0

°C

 

 

 

0.05 °C

 

 

 

 

 

0

°C

to

100

°C

 

 

 

0.08 °C

 

 

 

 

 

100

°C

to

250

°C

 

 

 

0.10 °C

 

 

 

 

 

250

°C

to

350

°C

 

 

 

0.43 °C

Comparison with a standard resistance sensor in dry block.

 

 

 

 

350

°C

to

600

°C

 

 

 

0.62 °C

 

 

 

3*

Glass thermometers

-40

°C

to

0

°C

 

 

 

0.20 °C

Comparison with a standard resistance sensor in liquid bath.

KP 3/T

1, 4

 

 

 

 

 

0

°C

 

 

 

0.05 °C

 

 

 

 

 

0

°C

to

100

°C

 

 

 

0.14 °C

 

 

 

 

 

100

°C

to

250

°C

 

 

 

0.15 °C

 

 

 

 

 

250

°C

to

300

°C

 

 

 

0.30 °C

 

 

 

4*

Direct-indicating thermometers

-40

°C

to

0

°C

 

 

 

0.07 °C

Comparison with a standard resistance sensor in liquid bath.

KP 4/T

1, 4

 

 

 

 

 

0

°C

 

 

 

0.06 °C

 

 

 

 

 

0

°C

to

100

°C

 

 

 

0.09 °C

 

 

 

 

 

100

°C

to

250

°C

 

 

 

0.10 °C

 

 

 

 

 

250

°C

to

350

°C

 

 

 

0.44 °C

Comparison with a standard resistance sensor in dry block.

 

 

 

 

350

°C

to

600

°C

 

 

 

0.62 °C

 

 

 

 

 

600

°C

to

800

°C

 

 

 

1.6 °C

Comparison with a standard thermoelectric sensor in a horizontal oven

 

 

 

 

800

°C

to

1,200

°C

 

 

 

2.2 °C

 

 

 

5*

Controllers and indicators

0

°C

to

1,370

°C

Type K

 

 

0.3 °C

Simulation of thermoelectric temperature sensors using DC voltage

KP 5/T

1, 4

 

 

0

°C

to

1,780

°C

Type S

 

 

0.6 °C

 

 

 

 

 

200

°C

to

500

°C

Type B

 

 

1.5 °C

 

 

 

 

 

500

°C

to

1,820

°C

Type B

 

 

0.6 °C

 

 

 

 

 

0

°C

to

1,768

°C

Type R

 

 

0.6 °C

 

 

 

 

 

0

°C

to

1,200

°C

Type J

 

 

0.2 °C

 

 

 

 

 

0

°C

to

1,300

°C

Type N

 

 

0.2 °C

 

 

 

 

 

-50

°C

to

850

°C

RTD

 

 

0.05 °C

Simulation of resistance temperature sensors using DC electrical resistors

 

 

6*

Non-contact thermometers

-30

°C

to

23

°C

 

 

 

2.2 °C

Direct measurement on a black body, comparison with a standard contactless thermometer.

KP 6/T

1, 4

 

 

23

°C

to

150

°C

 

 

 

2.1 °C

 

 

 

 

 

150

°C

to

250

°C

 

 

 

2.2 °C

 

 

 

 

 

250

°C

to

350

°C

 

 

 

2.5 °C

 

 

 

 

 

350

°C

to

500

°C

 

 

 

3.0 °C

 

 

 

7*

Direct-indicating thermometers integrated in conditioning and thermal chambers and baths

-80

°C

to

250

°C

 

 

 

1.0 °C

Comparison with a standard digital thermometer in a chamber (characterisation)

KP 22/T

1, 4

 

 

250

°C

to

600

°C

 

 

 

1.1 °C

 

 

 

 

 

600

°C

to

700

°C

 

 

 

1.7 °C

 

 

 

 

 

700

°C

to

850

°C

 

 

 

1.9 °C

 

 

 

 

 

850

°C

to

1,200

°C

 

 

 

2.9 °C

 

 

 

8*

Direct-indicating thermometers and temperature measuring chains integrated in conditioning and thermal chambers and baths

-20

°C

to

250

°C

 

 

 

1.0 °C

Comparison with a standard digital thermometer in a chamber (characterisation according to the standard AMS 2750)

KP 21/T

1

 

 

250

°C

to

600

°C

 

 

 

1.1 °C

 

 

 

 

 

600

°C

to

700

°C

 

 

 

1.7 °C

 

 

 

 

 

700

°C

to

850

°C

 

 

 

1.9 °C

 

 

 

 

 

850

°C

to

1,000

°C

 

 

 

2.9 °C

 

 

 

9*

Temperature controllers and indicators of conditioning and thermal chambers and baths

-20

°C

to

1,370

°C

Type K

 

 

0.3 °C

Simulation of thermoelectric temperature sensors using DC voltage (characterisation according to the standard AMS 2750)

KP 21/T

1

 

 

-20

°C

to

1,300

°C

Type N

 

 

0.2 °C

 

 

 

 

 

0

°C

to

1,780

°C

Type S

 

 

0.6 °C

 

 

 

 

 

200

°C

to

500

°C

Type B

 

 

1.5 °C

 

 

 

 

 

500

°C

to

1,800

°C

Type B

 

 

0.6 °C

 

 

 

1)    Asterisk at the ordinal number identifies the calibrations, which the Laboratory is qualified to carry out outside the permanent laboratory premises.

2     The expanded measurement uncertainty is in accordance with ILAC-P14 and EA-4/02, part of CMC, and it is the lowest value of the respective uncertainty. If not stated otherwise, its coverage probability is approx. 95 %. If not stated otherwise, the uncertainty values stated without a unit are relative to the value measured. If the calibration is carried out outside the laboratory premises, the measurement uncertainty may be affected.

3     If the document identifying the calibration procedure is dated, only these specific procedures are used. If the document identifying the calibration procedure is not dated, the latest edition of the specified procedure is used (including any changes).

 

 

CMC for the field of measured quantity:  Electrical quantities

Ord. num-ber1

Calibrated quantity / Subject of calibration

Nominal range

Parameter(s) of the meas. quantity

Lowest expanded measurement uncertainty specified2

Calibration principle

Calibration procedure identification3

Work-place

 

min.

unit

 

max.

unit

 

1

Direct-current voltage/ direct-current voltage meters (digital and analogue multimeters, voltmeters, oscilloscopes, testing and inspection instruments and their parts)

0

mV

to

1

mV

 

 

 

1.3 µV

Voltage generation by a standard calibrator

KP 1/E, KP 2/E, KP 3/E, KP 18/E

1

 

 

 

1

mV

to

20

mV

 

 

 

1.7 µV

 

 

 

 

 

 

20

mV

to

100

mV

 

 

 

2.1 µV

 

 

 

 

 

 

100

mV

to

200

mV

 

 

 

3.4 µV

 

 

 

 

 

 

200

mV

to

500

mV

 

 

 

0.5 µV

 

 

 

 

 

 

500

mV

to

900

mV

 

 

0.0011 %

 

 

 

 

 

 

 

0.9

V

to

1.5

V

 

 

0.0009 %

 

 

 

 

 

 

 

1.5

V

to

2

V

 

 

0.0008 %

 

 

 

 

 

 

 

2

V

to

4

V

 

 

0.0007 %

 

 

 

 

 

 

 

4

V

to

10

V

 

 

0.00055 %

 

 

 

 

 

 

 

10

V

to

20

V

 

 

0.00051 %

 

 

 

 

 

 

 

20

V

to

40

V

 

 

0.0011 %

 

 

 

 

 

 

 

40

V

to

100

V

 

 

0.00094 %

 

 

 

 

 

 

 

100

V

to

200

V

 

 

0.00082 %

 

 

 

 

 

 

 

200

V

to

400

V

 

 

0.0014 %

 

 

 

 

 

 

 

400

V

to

1,100

V

 

 

0.0013 %

 

 

 

 

 

 

 

1,100

V

to

6,000

V

 

 

0.6 %

 

 

 

 

 

2

Direct-current voltage / direct-current voltage sources and calibrators, multifunction calibrators, testing and inspection instruments and their parts, high voltage testers.

0

mV

to

120

mV

 

 

0.00096 %

+ 0.71 mV

Direct measurement by a standard multimeter

KP 8/E

1

 

 

 

120

mV

to

1,200

mV

 

 

0.00072 %

+ 4.6 mV

 

 

 

 

 

 

1.2

V

to

12

V

 

 

0.00060 %

+ 4.6 mV

 

 

 

 

 

 

12

V

to

120

V

 

 

0.00088 %

+ 29 mV

 

 

 

 

 

 

120

V

to

1,100

V

 

 

0.00098 %

+ 82 mV

 

 

 

 

 

 

1,100

V

to

6,000

V

 

 

0.6 %

 

Measurement by a standard multimeter with a HV probe

 

 

 

3

Direct current / direct-current meters (digital and analogue multimeters, ammeters, testing and inspection instruments and their parts, current probes).

0

mA

to

2

mA

 

 

 

3.8 nA

Current generation by a standard calibrator

KP 1/E, KP 2/E, KP 3/E

1

 

 

 

2

mA

to

10

mA

 

 

 

4.5 nA

 

 

 

 

 

 

10

mA

to

20

mA

 

 

 

5.5 nA

 

 

 

 

 

 

20

mA

to

40

mA

 

 

0.027 %

 

 

 

 

 

 

 

40

mA

to

100

mA

 

 

0.019 %

 

 

 

 

 

 

 

100

mA

to

200

mA

 

 

0.015 %

 

 

 

 

 

 

 

200

mA

to

200

mA

 

 

0.011 %

 

 

 

 

 

 

 

200

mA

to

400

mA

 

 

0.027 %

 

 

 

 

 

 

 

400

mA

to

1

A

 

 

0.019 %

 

 

 

 

 

 

 

1

A

to

2

A

 

 

0.015 %

 

 

 

 

 

 

 

2

A

to

3

A

 

 

0.045 %

 

 

 

 

 

 

 

3

A

to

4

A

 

 

0.038 %

 

Comparison with current measured by a standard shunt with multimeter

 

 

 

 

 

4

A

to

30

A

 

 

0.035 %

 

 

 

 

 

 

 

30

A

to

50

A

 

 

0.15 %

 

 

 

 

 

 

 

50

A

to

100

A

 

 

0.28 %

 

 

 

 

 

 

Direct current/ clamp ammeters and multimeters, clamp current probes.

1

A

to

5

A

 

 

0.2 %

 

Current generation by a standard calibrator

 

 

 

 

 

5

A

to

750

A

 

 

0.13 %

 

Simulation of reference current using calibrator and current coil

 

 

 

4

Direct current / current sources, testing and inspection instruments and their parts.

0

mA

to

12

mA

 

 

0.0045 %

+ 6.0 nA

Direct measurement by a standard multimeter

KP 1/E, KP 2/E, KP 3/E

1

 

 

 

12

mA

to

120

mA

 

 

0.0050 %

+ 6.7 nA

 

 

 

 

 

 

120

mA

to

1.2

mA

 

 

0.0050 %

+ 4.5 nA

 

 

 

 

 

 

1.2

mA

to

12

mA

 

 

0.0044 %

+ 0.045 mA

 

 

 

 

 

 

12

mA

to

120

mA

 

 

0.0065 %

+ 0.45 mA

 

 

 

 

 

 

120

mA

to

1.05

A

 

 

0.016 %

+ 9.8 mA

 

 

 

 

 

 

1.05

A

to

12

A

 

 

0.0073 %

+ 25 mA

Indirect measurement using a standard shunt and multimeter

KP 8/E

 

 

 

 

12

A

to

20

A

 

 

0.0058 %

+ 4 mA

 

 

 

 

 

 

20

A

to

50

A

 

 

0.042 %

+ 61 mA

 

 

 

 

 

 

50

A

to

100

A

 

 

0.047 %

 

 

 

 

 

 

Direct current / current sources

1

A

to

40

A

 

 

3.8 %

 

Direct measurement by a standard clamp multimeter

KP 8/E

1

 

 

 

40

A

to

100

A

 

 

5 %

 

 

 

 

 

 

 

100

A

to

250

A

 

 

4 %

 

 

 

 

 

 

 

250

A

to

400

A

 

 

2.5 %

 

 

 

 

 

 

 

400

A

to

1,000

A

 

 

2 %

 

 

 

 

 

5

Alternating-current voltage/ analogue multimeters, voltmeters, oscilloscopes, testing and inspection instruments and their parts

0.1

mV

to

1.0

mV

10 Hz to 100 kHz

 

 

8 mV

Voltage generation by a standard calibrator

KP 1/E, KP 2/E, KP 3/E, KP 18/E

1

 

 

 

 

 

 

 

 

100 kHz to 300 kHz

 

 

18 mV

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

 

34 mV

 

 

 

 

 

 

1.0

mV

to

10.0

mV

10 Hz to 100 kHz

 

 

15 mV

 

 

 

 

 

 

 

 

 

 

 

100 kHz to 300 kHz

 

 

30 mV

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

 

70 mV

 

 

 

 

 

 

10.0

mV

to

20.0

mV

10 Hz to 30 kHz

 

 

13mV

 

 

 

 

 

 

 

 

 

 

 

30 kHz to 100 kHz

 

 

18 mV

 

 

 

 

 

 

 

 

 

 

 

100 kHz to 300 kHz

 

 

45 mV

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

 

110 mV

 

 

 

 

 

 

20.0

mV

to

40.0

mV

10 Hz to 30 kHz

 

 

17 mV

 

 

 

 

 

 

 

 

 

 

 

30 kHz to 100 kHz

 

 

29 mV

 

 

 

 

 

 

 

 

 

 

 

100 kHz to 300 kHz

 

 

75 mV

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

 

270 mV

 

 

 

 

 

 

40.0

mV

to

200.0

mV

10 Hz to 30 kHz

 

0.045 %

 

 

 

 

 

 

 

 

 

 

 

 

30 kHz to 100 kHz

 

0.075 %

 

 

 

 

 

 

 

 

 

 

 

 

100 kHz to 300 kHz

 

0.2 %

 

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

0.7 %

 

 

 

 

 

 

 

200.0

mV

to

400.0

mV

10 Hz to 300 Hz

 

0.04 %

 

 

 

 

 

 

 

 

 

 

 

 

300 Hz to 30 kHz

 

0.03 %

 

 

 

 

 

 

 

 

 

 

 

 

30 kHz to 100 kHz

 

0.04 %

 

 

 

 

 

 

 

 

 

 

 

 

100 kHz to 300 kHz

 

0.1 %

 

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

0.55 %

 

 

 

 

 

 

 

0.4

V

to

1

V

10 Hz to 300 Hz

 

0.022 %

 

 

 

 

 

 

 

 

 

 

 

 

300 Hz to 30 kHz

 

0.016 %

 

 

 

 

 

 

 

 

 

 

 

 

30 kHz to 100 kHz

 

0.025 %

 

 

 

 

 

 

 

 

 

 

 

 

100 kHz to 300 kHz

 

0.07 %

 

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

0.47 %

 

 

 

 

 

 

 

1

V

to

2

V

10 Hz to 300 Hz

 

0.015 %

 

 

 

 

 

 

 

 

 

 

 

 

300 Hz to 30 kHz

 

0.01 %

 

 

 

 

 

 

 

 

 

 

 

 

30 kHz to 100 kHz

 

0.02 %

 

 

 

 

 

 

 

 

 

 

 

 

100 kHz to 300 kHz

 

0.05 %

 

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

0.43 %

 

 

 

 

 

 

 

2

V

to

4

V

10 Hz to 300 Hz

 

0.03 %

 

 

 

 

 

 

 

 

 

 

 

 

300 Hz to 30 kHz

 

0.02 %

 

 

 

 

 

 

 

 

 

 

 

 

30 kHz to 100 kHz

 

0.03 %

 

 

 

 

 

 

 

 

 

 

 

 

100 kHz to 300 kHz

 

0.11 %

 

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

0.55 %

 

 

 

 

 

 

 

4

V

to

10

V

10 Hz to 300 Hz

 

0.02 %

 

 

 

 

 

 

 

 

 

 

 

 

300 Hz to 30 kHz

 

0.013 %

 

 

 

 

 

 

 

 

 

 

 

 

30 kHz to 100 kHz

 

0.02 %

 

 

 

 

 

 

 

 

 

 

 

 

100 kHz to 300 kHz

 

0.086 %

 

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

0.47 %

 

 

 

 

 

 

 

10

V

to

20

V

10 Hz to 300 Hz

 

0.015 %

 

 

 

 

 

 

 

 

 

 

 

 

300 Hz to 30 kHz

 

0.01 %

 

 

 

 

 

 

 

 

 

 

 

 

30 kHz to 100 kHz

 

0.015 %

 

 

 

 

 

 

 

 

 

 

 

 

100 kHz to 300 kHz

 

0.075 %

 

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

0.42 %

 

 

 

 

 

 

 

20

V

to

40

V

10 Hz to 300 Hz

 

0.03 %

 

 

 

 

 

 

 

 

 

 

 

 

300 Hz to 30 kHz

 

0.019 %

 

 

 

 

 

 

 

 

 

 

 

 

30 kHz to 100 kHz

 

0.035 %

 

 

 

 

 

 

 

 

 

 

 

 

100 kHz to 300 kHz

 

0.11 %

 

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

0.15 %

 

 

 

 

 

 

 

40

V

to

100

V

10 Hz to 300 Hz

 

0.022 %

 

 

 

 

 

 

 

 

 

 

 

 

300 Hz to 10 kHz

 

0.01 %

 

 

 

 

 

 

 

 

 

 

 

 

10 kHz to 30 kHz

 

0.014 %

 

 

 

 

 

 

 

 

 

 

 

 

30 kHz to 100 kHz

 

0.03 %

 

 

 

 

 

 

 

 

 

 

 

 

100 kHz to 200 kHz

 

0.08 %

 

 

 

 

 

 

 

100

V

to

200

V

10 Hz to 300 Hz

 

0.02 %

 

 

 

 

 

 

 

 

 

 

 

 

300 Hz to 30 kHz

 

0.01 %

 

 

 

 

 

 

 

 

 

 

 

 

30 kHz to 100 kHz

 

0.023 %

 

 

 

 

 

 

 

 

 

 

 

 

100 kHz to 200 kHz

 

0.065 %

 

 

 

 

 

 

 

200

V

to

750

V

45 Hz to 300 Hz

 

0.03 %

 

 

 

 

 

 

 

 

 

 

 

 

300 Hz to 10 kHz

 

0.025 %

 

 

 

 

 

 

 

 

 

 

 

 

10 kHz to 30 kHz

 

0.03 %

 

 

 

 

 

 

 

 

 

 

 

 

30 kHz to 100 kHz

 

0.15 %

 

 

 

 

 

 

 

750

V

to

1,100

V

45 Hz to 300 Hz

 

0.027 %

 

 

 

 

 

 

 

 

 

 

 

 

300 Hz to 10 kHz

 

0.022 %

 

 

 

 

 

 

 

 

 

 

 

 

10 kHz to 30 kHz

 

0.025 %

 

 

 

 

 

 

 

1,100

V

to

5,000

V

40 Hz to 100 Hz

 

2 %

 

Comparison with a standard multimeter with a HV probe

 

 

 

6

Alternating-current voltage / voltage sources and calibrators, multifunction calibrators, testing and inspection instruments and their parts, high voltage testers

0.1

mV

to

12

mV

1 Hz to 40 Hz

 

0.062 %

+ 3,5 mV

Direct measurement by a standard multimeter

KP 8/E

1

 

 

 

 

 

 

 

 

40 Hz to 1,000 Hz

 

0.048 %

+ 1,4 mV

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 20 kHz

 

0.048 %

+ 1,4 mV

 

 

 

 

 

 

 

 

 

 

 

20 kHz to 50 kHz

 

0.13 %

+ 1,4 mV

 

 

 

 

 

 

 

 

 

 

 

50 kHz to 100 kHz

 

0.6 %

+ 1,4 mV

 

 

 

 

 

 

 

 

 

 

 

100 kHz to 300 kHz

 

4.7 %

+ 2,4 mV

 

 

 

 

 

 

12

mV

to

120

mV

1 Hz to 40 Hz

 

0.036 %

+ 2,8 mV

 

 

 

 

 

 

 

 

 

 

 

40 Hz to 1,000 Hz

 

0.017 %

+ 1.7 mV

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 20 kHz

 

0.017 %

+ 1.7 mV

 

 

 

 

 

 

 

 

 

 

 

20 kHz to 50 kHz

 

0.038 %

+ 1.7 mV

 

 

 

 

 

 

 

 

 

 

 

50 kHz to 100 kHz

 

0.13 %

+ 1.9 mV

 

 

 

 

 

 

 

 

 

 

 

100 kHz to 300 kHz

 

0.37 %

+ 12 mV

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

1.2 %

+ 11 mV

 

 

 

 

 

 

120

mV

to

1.2

V

1 Hz to 40 Hz

 

0.028 %

+ 32 mV

 

 

 

 

 

 

 

 

 

 

 

40 Hz to 1,000 Hz

 

0.01 %

+ 22 mV

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 20 kHz

 

0.01 %

+ 22 mV

 

 

 

 

 

 

 

 

 

 

 

20 kHz to 50 kHz

 

0.035 %

+ 22 mV

 

 

 

 

 

 

 

 

 

 

 

50 kHz to 100 kHz

 

0.093 %

+ 23 mV

 

 

 

 

 

 

 

 

 

 

 

100 kHz to 300 kHz

 

0.35 %

+ 120 mV

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

1.2 %

+ 80 mV

 

 

 

 

 

 

1.2

V

to

12

V

1 Hz to 40 Hz

 

0.028 %

+ 320 mV

 

 

 

 

 

 

 

 

 

 

 

40 Hz to 1,000 Hz

 

0.0093 %

+ 220 mV

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 20 kHz

 

0.0093 %

+ 220 mV

 

 

 

 

 

 

 

 

 

 

 

20 kHz to 50 kHz

 

0.035 %

+ 220 mV

 

 

 

 

 

 

 

 

 

 

 

50 kHz to 100 kHz

 

0.093 %

+ 230 mV

 

 

 

 

 

 

 

 

 

 

 

100 kHz to 300 kHz

 

0.35 %

+ 1.2 mV

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

1.2 %

+ 0.77 mV

 

 

 

 

 

 

12

V

to

120

V

1 Hz to 40 Hz

 

0.036 %

+ 3.8 mV

 

 

 

 

 

 

 

 

 

 

 

40 Hz to 1,000 Hz

 

0.024 %

+ 2.2 mV

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 20 kHz

 

0.024 %

+ 2.2 mV

 

 

 

 

 

 

 

 

 

 

 

20 kHz to 50 kHz

 

0.041 %

+ 2.2 mV

 

 

 

 

 

 

 

 

 

 

 

50 kHz to 100 kHz

 

0.14 %

+ 2.3 mV

 

 

 

 

 

 

 

 

 

 

 

100 kHz to 300 kHz

 

0.47 %

+ 12 mV

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

1.2 %

+ 16 mV

 

 

 

 

 

 

120

V

to

700

V

1 Hz to 40 Hz

 

0.047 %

+ 46 mV

 

 

 

 

 

 

 

 

 

 

 

40 Hz to 1,000 Hz

 

0.047 %

+ 23 mV

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 20 kHz

 

0.047 %

+ 23 mV

 

 

 

 

 

 

 

 

 

 

 

20 kHz to 50 kHz

 

0.014 %

+ 23 mV

 

 

 

 

 

 

 

 

 

 

 

50 kHz to 100 kHz

 

0.24 %

+ 23 mV

 

 

 

 

 

 

1,100

V

to

5,000

V

40 Hz to 100 Hz

 

2 %

 

Measurement by a standard multimeter with a HV probe

KP 8/E

1

 

7

Alternating-current voltage, peak and peak-to-peak value of periodic non-harmonic signals / sources of non-harmonic and pulse signals, pulse testers

 

 

 

 

 

Bandwidth:

Impedance:

 

 

Measurement by a standard oscilloscope5

KP 8/E

1

 

 

 

10

mV

to

5

V

1 Hz to 300 kHz,

Z = 50 W

1.8 %

 

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 2 GHz

 

4.0 %

 

 

 

 

 

 

 

5

V

to

38

V

1 Hz to 300 kHz,

Z = 50 W

2.9 %

 

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 GHz

 

4.5 %

 

 

 

 

 

 

 

10

mV

to

40

V

1 Hz to 300 kHz,

Z = 1 MW

1.8 %

 

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 400 MHz

 

4.0 %

 

 

 

 

 

 

 

40

V

to

400

V

1 Hz to 400 MHz

Z = 10 MW

4.5 %

 

 

 

 

 

 

 

40

V

to

1,000

V

1 Hz to 400 MHz

Z = 66.7 MW

4.5 %

 

 

 

 

 

8

Pulse amplitude, peak and peak-to-peak value of voltage / sources of non-harmonic and pulse signals, pulse testers

10

mV

to

40

V

Z = 1 MW

rise time >1.3 ns

4.0 %

 

Measurement by a standard oscilloscope with a voltage probe

KP 8/E

1

 

 

 

10

V

to

5,000

V

Z = 50 kW

rise time >140 ns

5.6 %

 

 

 

 

 

 

 

10

V

to

5,000

V

Z = 1 kW

rise time >4.2 ns

5.0 %

 

 

 

 

 

 

 

1

V

to

5,000

V

Z = 50 W

rise time >3.5 ns

4.9 %

 

 

 

 

 

9

Alternating current, peak and peak-to-peak value / sources of non-harmonic and pulse signals, pulse testers

100

mA

to

1.5

A

1 Hz to 20 kHz

 

2.7 %

 

Measurement by a standard oscilloscope with a shunt or clamp probe

KP 8/E

1

 

 

 

1.5

A

to

15

A

1 Hz to 20 kHz

 

2.4 %

 

 

 

 

 

 

 

1

A

to

2,500

A

 

rise time >140 ns

4.9 %

 

Measurement by a standard oscilloscope with a current shunt

 

 

 

10

Alternating current / digital, analogue multimeters and ammeters, testing and inspection instruments and their parts, current probes

1

mA

to

10

mA

10 Hz to 1 kHz

 

 

0.015 mA

Current generation by a standard calibrator

KP 1/E, KP 2/E, KP 3/E

1

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

 

0.02 mA

 

 

 

 

 

 

10

mA

to

50

mA

10 Hz to 1 kHz

 

 

0.025 mA

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

 

0.035 mA

 

 

 

 

 

 

50

mA

to

100

mA

10 Hz to 1 kHz

 

 

0.035 mA

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

 

0.06 mA

 

 

 

 

 

 

100

mA

to

200

mA

10 Hz to 1 kHz

 

0.035 %

 

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

0.06 %

 

 

 

 

 

 

 

200

mA

to

400

mA

10 Hz to 1 kHz

 

0.071 %

 

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

0.083 %

 

 

 

 

 

 

 

400

mA

to

800

mA

10 Hz to 1 kHz

 

0.045 %

 

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

0.06 %

 

 

 

 

 

 

 

800

mA

to

2

mA

10 Hz to 1 kHz

 

0.03 %

 

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

0.045 %

 

 

 

 

 

 

 

2

mA

to

3

mA

10 Hz to 1 kHz

 

0.075 %

 

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

0.085 %

 

 

 

 

 

 

 

3

mA

to

4

mA

10 Hz to 1 kHz

 

0.055 %

 

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

0.065 %

 

 

 

 

 

 

 

4

mA

to

8

mA

10 Hz to 1 kHz

 

0.045 %

 

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

0.06 %

 

 

 

 

 

 

 

8

mA

to

20

mA

10 Hz to 1 kHz

 

0.033 %

 

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

0.045 %

 

 

 

 

 

 

 

20

mA

to

30

mA

10 Hz to 1 kHz

 

0.075 %

 

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

0.085 %

 

 

 

 

 

 

 

30

mA

to

40

mA

10 Hz to 1 kHz

 

0.055 %

 

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

0.065 %

 

 

 

 

 

 

 

40

mA

to

80

mA

10 Hz to 1 kHz

 

0.045 %

 

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

0.055 %

 

 

 

 

 

 

 

80

mA

to

200

mA

10 Hz to 1 kHz

 

0.03 %

 

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

0.045 %

 

 

 

 

 

 

 

200

mA

to

300

mA

10 Hz to 1 kHz

 

0.15 %

 

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

2 %

 

 

 

 

 

 

 

300

mA

to

800

mA

10 Hz to 1 kHz

 

0.1 %

 

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

0.15 %

 

 

 

 

 

 

 

800

mA

to

2,000

mA

10 Hz to 1 kHz

 

0.06 %

 

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

0.11 %

 

 

 

 

 

 

 

2

A

to

4

A

10 Hz to 1 kHz

 

0.1 %

 

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

0.2 %

 

 

 

 

 

 

 

 

 

 

 

 

5 kHz to 10 kHz

 

0.5 %

 

 

 

 

 

 

 

 

 

 

 

 

10 kHz to 20 kHz

 

2 %

 

 

 

 

 

 

 

4

A

to

6

A

10 Hz to 1 kHz

 

0.07 %

 

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

0.14 %

 

 

 

 

 

 

 

 

 

 

 

 

5 kHz to 10 kHz

 

0.38 %

 

 

 

 

 

 

 

 

 

 

 

 

10 kHz to 20 kHz

 

1.5 %

 

 

 

 

 

 

 

6

A

to

11

A

10 Hz to 1 kHz

 

0.07 %

 

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

0.13 %

 

 

 

 

 

 

 

 

 

 

 

 

5 kHz to 10 kHz

 

0.35 %

 

 

 

 

 

 

 

 

 

 

 

 

10 kHz to 20 kHz

 

1.3 %

 

 

 

 

 

 

 

11

A

to

20

A

40 Hz to 400 Hz

 

0.35 %

 

 

 

 

 

 

 

20

A

to

30

A

40 Hz to 300 Hz

 

0.1 %

 

Comparison with current measured by a standard shunt with multimeter

KP 1/E, KP 2/E, KP  3/E

1

 

 

 

30

A

to

50

A

40 Hz to 250 Hz

 

1.2 %

 

 

 

 

 

 

 

50

A

to

100

A

40 Hz to 160 Hz

 

2 %

 

 

 

 

 

 

Clamp ammeters and clamp current probes

1

A

to

5

A

40 Hz to 300 Hz

 

0.25 %

 

Direct generation of current with a calibrator

KP 1/E

 

 

 

 

5

A

to

20

A

40 Hz to 300 Hz

 

0.16 %

 

Simulation of reference current using calibrator and current coil

 

 

 

 

 

20

A

to

30

A

40 Hz to 300 Hz

 

0.33 %

 

 

 

 

 

 

 

30

A

to

100

A

40 Hz to 300 Hz

 

0.25 %

 

 

 

 

 

 

 

100

A

to

250

A

40 Hz to 300 Hz

 

0.35 %

 

 

 

 

 

 

 

250

A

to

750

A

40 Hz to 300 Hz

 

0.2 %

 

 

 

 

 

11

Alternating current / current sources

1

mA

to

120

mA

10 Hz to 20 Hz

 

0.46 %

+ 35 nA

Direct measurement by a standard multimeter

KP 8/E

1

 

 

 

 

 

 

 

 

20 Hz to 45 Hz

 

0.17 %

+ 35 nA

 

 

 

 

 

 

 

 

 

 

 

45 Hz to 100 Hz

 

0.07 %

+ 35 nA

 

 

 

 

 

 

 

 

 

 

 

100 Hz to 1,000 Hz

 

0.071 %

+ 35 nA

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

0.061 %

+ 35 nA

 

 

 

 

 

 

 

 

 

 

 

5 kHz to 20 kHz

 

0.061 %

+ 35 nA

 

 

 

 

 

 

120

mA

to

1.2

mA

10 Hz to 20 Hz

 

0.46 %

+ 230 nA

 

 

 

 

 

 

 

 

 

 

 

20 Hz to 45 Hz

 

0.17 %

+ 230 nA

 

 

 

 

 

 

 

 

 

 

 

45 Hz to 100 Hz

 

0.07 %

+ 230 nA

 

 

 

 

 

 

 

 

 

 

 

100 Hz to 1,000 Hz

 

0.037 %

+ 230 nA

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

0.044 %

+ 230 nA

 

 

 

 

 

 

 

 

 

 

 

5 kHz to 20 kHz

 

0.044 %

+ 230 nA

 

 

 

 

 

 

1.2

mA

to

12

mA

10 Hz to 20 Hz

 

0.46 %

+ 2.3 mA

 

 

 

 

 

 

 

 

 

 

 

20 Hz to 45 Hz

 

0.17 %

+ 2.3 mA

 

 

 

 

 

 

 

 

 

 

 

45 Hz to 100 Hz

 

0.07 %

+ 2.3 mA

 

 

 

 

 

 

 

 

 

 

 

100 Hz to 1,000 Hz

 

0.037 %

+ 2.3 mA

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

0.041 %

+ 2.3 mA

 

 

 

 

 

 

 

 

 

 

 

5 kHz to 20 kHz

 

0.041 %

+ 2.3 mA

 

 

 

 

 

 

12

mA

to

120

mA

10 Hz to 20 Hz

 

0.46 %

+ 23 mA

 

 

 

 

 

 

 

 

 

 

 

20 Hz to 45 Hz

 

0.17 %

+ 23 mA

 

 

 

 

 

 

 

 

 

 

 

45 Hz to 100 Hz

 

0.07 %

+ 23 mA

 

 

 

 

 

 

 

 

 

 

 

100 Hz to 1,000 Hz

 

0.037 %

+ 23 mA

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

0.04 %

+ 23 mA

 

 

 

 

 

 

 

 

 

 

 

5 kHz to 20 kHz

 

0.04 %

+ 23 mA

 

 

 

 

 

 

120

mA

to

1

A

10 Hz to 20 Hz

 

0.46 %

+ 230 mA

 

 

 

 

 

 

 

 

 

 

 

20 Hz to 45 Hz

 

0.17 %

+ 230 mA

 

 

 

 

 

 

 

 

 

 

 

45 Hz to 100 Hz

 

0.07 %

+ 230 mA

 

 

 

 

 

 

 

 

 

 

 

100 Hz to 1,000 Hz

 

0.12 %

+ 230 mA

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 5 kHz

 

0.13 %

+ 230 mA

 

 

 

 

 

 

 

 

 

 

 

5 kHz to 20 kHz

 

0.13 %

+ 230 mA

 

 

 

 

 

 

1.2

A

to

12

A

40 Hz to 1 kHz

 

0.14 %

+ 330 mA

Indirect measurement by a multimeter with a standard shunt

KP 8/E

1

 

 

 

 

 

 

 

 

1 kHz to 20 kHz

 

0.15 %

+ 41 mA

 

 

 

 

 

 

12

A

to

20

A

40 Hz to 1 kHz

 

0.14 %

+ 420 mA

 

 

 

 

 

 

 

 

 

 

 

1 kHz to 20 kHz

 

0.15 %

+ 490 mA

 

 

 

 

 

 

20

A

to

50

A

40 Hz to 250 Hz

 

1 %

+ 44 mA

 

 

 

 

 

 

50

A

to

100

A

40 Hz to 250 Hz

 

2 %

+ 48 mA

 

 

 

 

 

 

1

A

to

40

A

40 Hz to 300 Hz

 

1.2 %

+ 0.5 A

Measurement by a standard clamp ammeter

KP 8/E

1

 

 

 

40

A

to

400

A

40 Hz to 300 Hz

 

0.6 %

+ 5 A

 

 

 

 

 

 

400

A

to

1,500

A

40 Hz to 150 Hz

 

0.6 %

+ 23 A

 

 

 

 

12*

DC resistance / ohmmeters, testing and inspection instruments and their parts.

 

 

 

0

W

 

 

 

20 mW

Direct measurement of fixed resistance standards

KP 1/E, KP 2/E

1

 

 

 

 

 

 

0.1

mW

 

 

 

6 nW

 

 

 

 

 

 

 

 

 

1

mW

 

 

 

50 nW

 

 

 

 

 

 

 

 

 

0.01

W

 

 

 

0.3 mW

 

 

 

 

 

 

 

 

 

0.1

W

 

 

 

2 mW

 

 

 

 

 

 

 

 

 

1

W

 

 

 

12 mW

 

 

 

 

 

 

 

 

 

10

W

 

 

 

200 mW

 

 

 

 

 

 

 

 

 

100

W

 

 

 

1.2 mW

 

 

 

 

 

 

 

 

 

1

kW

 

 

 

14 mW

 

 

 

 

 

 

 

 

 

10

kW

 

 

 

0.14 W

 

 

 

 

 

 

 

 

 

100

kW

 

 

 

1.7 W

 

 

 

 

 

 

 

 

 

1

MW

 

 

 

30 W

 

 

 

 

 

 

 

 

 

10

MW

 

 

 

60 W

 

 

 

 

 

 

 

 

 

100

MW

 

 

 

9 kW

 

 

 

 

 

 

 

 

 

1

GW

 

 

 

0.4 MW

 

 

 

 

 

 

 

 

 

10

GW

 

 

 

0.4 GW

 

 

 

 

 

 

 

 

 

100

GW

 

 

 

5 GW

 

 

 

 

 

 

 

 

 

1

TW

 

 

 

100 GW

 

 

 

 

13*

DC resistance/ standard resistors, shunts

 

 

 

0.1

mW

 

 

 

10 nW

Comparison by ratio method – measurement of voltage drop at constant measuring current

KP 6/E

1

 

 

 

 

 

 

1

mW

 

 

 

70 nW

 

 

 

 

 

 

 

 

 

0.01

W

 

 

 

0.4 mW

 

 

 

 

 

 

 

 

 

0.1

W

 

 

 

3 mW

 

 

 

 

 

 

 

 

 

1

W

 

 

 

20 mW

 

 

 

 

 

 

 

 

 

10

W

 

 

 

250 mW

 

 

 

 

 

 

 

 

 

100

W

 

 

 

1.5 mW

 

 

 

 

 

 

 

 

 

1

kW

 

 

 

40 mW

 

 

 

 

 

 

 

 

 

10

kW

 

 

 

0.4 W

 

 

 

 

 

 

 

 

 

100

kW

 

 

 

7 W

 

 

 

 

 

 

 

 

 

1

MW

 

 

 

40 W

Comparison by ratio method – measurement of standard and calibrated resistance

 

 

 

 

 

 

 

 

10

MW

 

 

 

700 W

 

 

 

 

 

 

 

 

 

100

MW

 

 

 

12 kW

 

 

 

 

 

 

 

 

 

1

GW

 

 

 

0.6 MW

 

 

 

 

14*

DC resistance / resistors, calibrators, simulators, testing and inspection instruments and their parts.

0

W

to

12

W

 

 

0.002 %

+ 78 mW

Direct measurement by a standard multimeter

KP 8/E

1

 

 

 

12

W

to

120

W

 

 

0.0016 %

+ 570 mW

 

 

 

 

 

 

0.12

kW

to

1.2

kW

 

 

0.0014 %

+ 540 mW

 

 

 

 

 

 

1.2

kW

to

12

kW

 

 

0.0014 %

+ 5.4 mW

 

 

 

 

 

 

12

kW

to

120

kW

 

 

0.0016 %

+ 45 mW

 

 

 

 

 

 

0.12

MW

to

1.2

MW

 

 

0.0026 %

+ 1.9 W

 

 

 

 

 

 

1.2

MW

to

12

MW

 

 

0.0068 %

+ 110 W

 

 

 

 

 

 

12

MW

to

120

MW

 

 

0.058 %

+ 1.3 kW

 

 

 

 

 

 

0.12

GW

to

1.2

GW

 

 

0.57 %

+ 0.35 MW

 

 

 

 

15

HF level (voltage) / HF voltmeters, measuring receivers, synthesizers, oscilloscopes, spectral and circuit analyzers, radiocommunication and radio navigation testers, simulators and imitators (ATC, DME, TACAN, TCAS, VOR-ILS)

20
(10

dBmV µV

to

47
224

dBmV mV)

0.1 MHz to 1.4 GHz

 

4 %

 

Indirect comparison with standard HF wattmeter

KP 13/E, KP 18/E

1

 

 

 

 

 

 

 

1.4 GHz to 2.08 GHz

 

3.5 %

 

 

 

 

 

 

47
(224

dBmV
µV

to

67
2.24

dBmV
mV)

10 MHz to 30 MHz

 

4 %

 

 

 

 

 

 

 

 

 

 

 

30 MHz to 2.08 GHz

 

3.5 %

 

 

 

 

 

 

67
(2.24

dBmV
µV

to

87
22.4

dBmV
mV)

10 MHz to 30 MHz

 

4 %

 

 

 

 

 

 

 

 

 

 

 

30 MHz to 2.08 GHz

 

3.5 %

 

 

 

 

 

 

87
(22.4

dBmV
mV

to

117
707

dBmV
mV)

100 kHz to 300 kHz

 

17 %

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 600 kHz

 

14 %

 

 

 

 

 

 

 

 

 

 

 

600 kHz to 1 MHz

 

8.5 %

 

 

 

 

 

 

 

 

 

 

 

1 MHz to 1.3 GHz

 

2.6 %

 

 

 

 

 

 

 

 

 

 

 

1.3 GHz to 2.08 GHz

 

3.2 %

 

 

 

 

 

 

117
(707

dBmV
mV

to

137
7.07

dBmV
V)

100 kHz to 300 kHz

 

17 %

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 600 kHz

 

14 %

 

 

 

 

 

 

 

 

 

 

 

600 kHz to 1 MHz

 

8.5 %

 

 

 

 

 

 

 

 

 

 

 

1 MHz to 1.3 GHz

 

3.5 %

 

 

 

 

 

 

 

 

 

 

 

1.3 GHz to 2.08 GHz

 

4 %

 

 

 

 

 

 

137
(7.07

dBmV
V

to

147
22.4

dBmV
V)

100 kHz to 2.08 GHz

 

2.9 %

 

 

 

 

 

 

147
(22.4

dBmV
V

to

152
39.0

dBmV
V)

100 kHz to 2.08 GHz

 

3.7 %

 

 

 

 

16

HF level (voltage) / HF generators, circuit analyzers, radiocommunication and radio navigation testers, simulators and imitators (ATC, DME, TACAN, TCAS, VOR-ILS)

20
(10

dBmV µV

to

47
224

dBmV mV)

0.1 MHz to 1.4 GHz

 

4 %

 

Indirect measurement of HF voltage - by measuring HF power at impedance 50 W

 

KP 11/E, KP 12/E,

1

 

 

 

 

 

 

 

1.4 GHz to 2.08 GHz

 

3.5 %

 

 

 

 

 

 

47
(224

dBmV
µV

to

67
2.24

dBmV
mV)

10 MHz to 30 MHz

 

4 %

 

 

 

 

 

 

 

 

 

 

 

30 MHz to 2.08 GHz

 

3.5 %

 

 

 

 

 

 

67
(2.24

dBmV
µV

to

87
22.4

dBmV
mV)

10 MHz to 30 MHz

 

4 %

 

 

 

 

 

 

 

 

 

 

 

30 MHz to 2.08 GHz

 

3.5 %

 

 

 

 

 

 

87
(22.4

dBmV
mV

to

117
707

dBmV
mV)

100 kHz to 300 kHz

 

17 %

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 600 kHz

 

14 %

 

 

 

 

 

 

 

 

 

 

 

600 kHz to 1 MHz

 

8.5 %

 

 

 

 

 

 

 

 

 

 

 

1 MHz to 1.3 GHz

 

2.6 %

 

 

 

 

 

 

 

 

 

 

 

1.3 GHz to 2.08 GHz

 

3.2 %

 

 

 

 

 

 

117
(707

dBmV
mV

to

137
7.07

dBmV
V)

100 kHz to 300 kHz

 

17 %

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 600 kHz

 

14 %

 

 

 

 

 

 

 

 

 

 

 

600 kHz to 1 MHz

 

8.5 %

 

 

 

 

 

 

 

 

 

 

 

1 MHz to 1.3 GHz

 

3.5 %

 

 

 

 

 

 

 

 

 

 

 

1.3 GHz to 2.08 GHz

 

4 %

 

 

 

 

 

 

137
(7.07

dBmV
V

to

147
22.4

dBmV
V)

100 kHz to 2.08 GHz

 

2.9 %

 

 

 

 

 

 

147
(22.4

dBmV
V

to

152
39.0

dBmV
V)

100 kHz to 2.08 GHz

 

3.7 %

 

 

 

 

17*

HF level (power) / oscillators, circuit analyzers, testers, radiocommunication and radio navigation testers, simulators and imitators (ATC, DME, TACAN, TCAS, VOR-ILS)

 

 

 

 

 

 

 

 

 

Direct measurement by a standard measuring receiver

KP 15/E

1

 

 

1
(-90

pW
dBm

to

2.5
-86

pW
dBm)

10 MHz to 20 MHz

 

8.7 %

 

 

 

 

 

 

 

 

 

 

 

20 MHz to 50 MHz

 

6.5 %

 

 

 

 

 

 

 

 

 

 

 

50 MHz to 7 GHz

 

6.2 %

 

 

 

 

 

 

2.5
(-86

pW
dBm

to

800
-61

pW
dBm)

10 MHz to 20 MHz

 

7.8 %

 

 

 

 

 

 

 

 

 

 

 

20 MHz to 50 MHz

 

5.2 %

 

 

 

 

 

 

 

 

 

 

 

50 MHz to 7 GHz

 

4.8 %

 

 

 

 

 

 

800
(-61

pW
dBm

to

2
-57

nW
dBm)

10 MHz to 20 MHz

 

8.5 %

 

Measurement by a standard HF wattmeter

KP 15/E

 

 

 

 

 

 

 

 

20 MHz to 50 MHz

 

6.2 %

 

 

 

 

 

 

 

 

 

 

 

50 MHz to 7 GHz

 

5.5 %

 

 

 

 

 

 

 

 

 

 

 

7 GHz to 15 GHz

 

6.0 %

 

 

 

 

 

 

 

 

 

 

 

15 GHz to 18 GHz

 

6.6 %

 

 

 

 

 

 

2
(-57

nW
dBm

to

10
-20

μW
dBm)

10 MHz to 20 MHz

 

7.5 %

 

 

 

 

 

 

 

 

 

 

 

20 MHz to 50 MHz

 

4.8 %

 

 

 

 

 

 

 

 

 

 

 

50 MHz to 7 GHz

 

3.9 %

 

 

 

 

 

 

 

 

 

 

 

7 GHz to 15 GHz

 

4.5 %

 

 

 

 

 

 

 

 

 

 

 

15 GHz to 18 GHz

 

5.4 %

 

 

 

 

 

 

10
(-20

μW
dBm

to

100
-10

μW
dBm)

100 kHz to 300 kHz

 

7.3 %

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

4.0 %

 

 

 

 

 

 

 

 

 

 

 

1 MHz to 50 MHz

 

3.6 %

 

 

 

 

 

 

 

 

 

 

 

50 MHz to 1 GHz

 

3.0 %

 

 

 

 

 

 

 

 

 

 

 

1 GHz to 2 GHz

 

3.7 %

 

 

 

 

 

 

 

 

 

 

 

2 GHz to 15 GHz

 

4.8 %

 

 

 

 

 

 

 

 

 

 

 

15 GHz to 18 GHz

 

5.1 %

 

 

 

 

 

 

 

 

 

 

 

18 GHz to 20 GHz

 

5.5 %

 

 

 

 

 

 

 

 

 

 

 

20 GHz to 26.5 GHz

 

4.6 %

 

 

 

 

 

 

 

 

 

 

 

26.5 GHz to 40.8 GHz

 

6.0 %

 

 

 

 

 

 

100
(-10

μW
dBm

to

10
+10

mW
dBm)

100 kHz to 300 kHz

 

7.7 %

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

4.1 %

 

 

 

 

 

 

 

 

 

 

 

1 MHz to 50 MHz

 

3.4 %

 

 

 

 

 

 

 

 

 

 

 

50 MHz to 1 GHz

 

2.6 %

 

 

 

 

 

 

 

 

 

 

 

1 GHz to 2 GHz

 

3.7 %

 

 

 

 

 

 

 

 

 

 

 

2 GHz to 15 GHz

 

4.6 %

 

 

 

 

 

 

 

 

 

 

 

15 GHz to 18 GHz

 

4.9 %

 

 

 

 

 

 

 

 

 

 

 

18 GHz to 20 GHz

 

5.3 %

 

 

 

 

 

 

 

 

 

 

 

20 GHz to 26.5 GHz

 

4.3 %

 

 

 

 

 

 

 

 

 

 

 

26.5 GHz to 40.8 GHz

 

5.6 %

 

 

 

 

 

 

10
(+10

mW
dBm

to

100
+20

mW
dBm)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

100 kHz to 300 kHz

 

6.2 %

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

6.2 %

 

 

 

 

 

 

 

 

 

 

 

1 MHz to 10 MHz

 

5.4 %

 

 

 

 

 

 

 

 

 

 

 

10 MHz to 50 MHz

 

4.6 %

 

 

 

 

 

 

 

 

 

 

 

50 MHz to 1 GHz

 

4.4 %

 

 

 

 

 

 

 

 

 

 

 

1 GHz to 15 GHz

 

5.8 %

 

 

 

 

 

 

 

 

 

 

 

15 GHz to 18 GHz

 

6.0 %

 

 

 

 

 

 

 

 

 

 

 

18 GHz to 20 GHz

 

6.4 %

 

 

 

 

 

 

 

 

 

 

 

20 GHz to 26.5 GHz

 

5.5 %

 

 

 

 

 

 

 

 

 

 

 

26.5 GHz to 40.8 GHz

 

6.5 %

 

 

 

 

 

 

100

mW

to

3

W

 

 

 

 

 

 

 

 

 

(+20

dBm

to

+33

dBm)

100 kHz to 300 kHz

 

6.0 %

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

6.0 %

 

 

 

 

 

 

 

 

 

 

 

1 MHz to 10 MHz

 

5.2 %

 

 

 

 

 

 

 

 

 

 

 

10 MHz to 500 MHz

 

4.8 %

 

 

 

 

 

 

 

 

 

 

 

500 MHz to 1 GHz

 

5.9 %

 

 

 

 

 

 

3
(+33

mW
dBm

to

10
+40

W
dBm)

1 GHz to 4.2 GHz

 

6.7 %

 

 

 

 

 

 

 

 

 

 

 

100 kHz to 300 kHz

 

7.3 %

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

7.3 %

 

 

 

 

 

 

 

 

 

 

 

1 MHz to 10 MHz

 

6.6 %

 

 

 

 

 

 

 

 

 

 

 

10 MHz to 500 MHz

 

6.3 %

 

 

 

 

 

 

 

 

 

 

 

500 MHz to 1 GHz

 

7.1 %

 

 

 

 

 

 

10
(+40

W
dBm

to

25
+44

W
dBm)

1 GHz to 4.2 GHz

 

7.6 %

 

 

 

 

 

 

 

 

 

 

 

100 kHz to 300 kHz

 

7.4 %

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

7.4 %

 

 

 

 

 

 

 

 

 

 

 

1 MHz to 10 MHz

 

6.7 %

 

 

 

 

 

 

 

 

 

 

 

10 MHz to 500 MHz

 

6.2 %

 

 

 

 

 

 

 

 

 

 

 

500 MHz to 1 GHz

 

6.9 %

 

 

 

 

 

 

 

 

 

 

 

1 GHz to 2 GHz

 

8.0 %

 

 

 

 

 

 

10
(+40

W
dBm

to

25
+44

W
dBm)

2 GHz to 4.2 GHz

 

9.5 %

 

 

 

 

 

 

 

 

 

 

 

1 GHz to 2 GHz

 

8.0 %

 

 

 

 

18*

HF level (power) / HF wattmeters, selective HF meters, spectral and circuit analyzers, testers, radiocommunication and radio navigation testers, simulators and imitators (ATC, DME, TACAN, TCAS, VOR-ILS)

 

 

 

 

 

 

 

 

 

Comparison with a standard HF wattmeter

KP 14/E

1

 

 

800
(-61

pW
dBm

to

2
-57

nW
dBm)

10 MHz to 20 MHz

 

8.5 %

 

 

 

 

 

 

 

 

 

 

 

20 MHz to 50 MHz

 

6.2 %

 

 

 

 

 

 

 

 

 

 

 

50 MHz to 7 GHz

 

5.5 %

 

 

 

 

 

 

 

 

 

 

 

7 GHz to 15 GHz

 

6.0 %

 

 

 

 

 

 

 

 

 

 

 

15 GHz to 18 GHz

 

6.6 %

 

 

 

 

 

 

2
(-57

nW
dBm

to

10
-20

μW
dBm)

10 MHz to 20 MHz

 

7.5 %

 

 

 

 

 

 

 

 

 

 

 

20 MHz to 50 MHz

 

4.8 %

 

 

 

 

 

 

 

 

 

 

 

50 MHz to 7 GHz

 

3.9 %

 

 

 

 

 

 

 

 

 

 

 

7 GHz to 15 GHz

 

4.5 %

 

 

 

 

 

 

 

 

 

 

 

15 GHz to 18 GHz

 

5.4 %

 

 

 

 

 

 

10
(-20

μW
dBm

to

100
-10

μW
dBm)

100 kHz to 300 kHz

 

7.3 %

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

4.0 %

 

 

 

 

 

 

 

 

 

 

 

1 MHz to 50 MHz

 

3.6 %

 

 

 

 

 

 

 

 

 

 

 

50 MHz to 1 GHz

 

3.0 %

 

 

 

 

 

 

 

 

 

 

 

1 GHz to 2 GHz

 

3.7 %

 

 

 

 

 

 

 

 

 

 

 

2 GHz to 15 GHz

 

4.8 %

 

 

 

 

 

 

 

 

 

 

 

15 GHz to 18 GHz

 

5.1 %

 

 

 

 

 

 

 

 

 

 

 

18 GHz to 20 GHz

 

5.5 %

 

 

 

 

 

 

 

 

 

 

 

20 GHz to 26.5 GHz

 

4.6 %

 

 

 

 

 

 

 

 

 

 

 

26.5 GHz to 40.8 GHz

 

6.0 %

 

 

 

 

 

 

100
(-10

μW
dBm

to

10
+10

mW
dBm)

100 kHz to 300 kHz

 

7.7 %

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

4.1 %

 

 

 

 

 

 

 

 

 

 

 

1 MHz to 50 MHz

 

3.4 %

 

 

 

 

 

 

 

 

 

 

 

50 MHz to 1 GHz

 

2.6 %

 

 

 

 

 

 

 

 

 

 

 

1 GHz to 2 GHz

 

3.7 %

 

 

 

 

 

 

 

 

 

 

 

2 GHz to 15 GHz

 

4.6 %

 

 

 

 

 

 

 

 

 

 

 

15 GHz to 18 GHz

 

4.9 %

 

 

 

 

 

 

 

 

 

 

 

18 GHz to 20 GHz

 

5.3 %

 

 

 

 

 

 

 

 

 

 

 

20 GHz to 26.5 GHz

 

4.3 %

 

 

 

 

 

 

 

 

 

 

 

26.5 GHz to 40.8 GHz

 

5.6 %

 

 

 

 

 

 

10
(+10

mW
dBm

to

100
+20

mW
dBm)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

100 kHz to 300 kHz

 

6.2 %

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

6.2 %

 

 

 

 

 

 

 

 

 

 

 

1 MHz to 10 MHz

 

5.4 %

 

 

 

 

 

 

 

 

 

 

 

10 MHz to 50 MHz

 

4.6 %

 

 

 

 

 

 

 

 

 

 

 

50 MHz to 1 GHz

 

4.4 %

 

 

 

 

 

 

 

 

 

 

 

1 GHz to 15 GHz

 

5.8 %

 

 

 

 

 

 

 

 

 

 

 

15 GHz to 18 GHz

 

6.0 %

 

 

 

 

 

 

 

 

 

 

 

18 GHz to 20 GHz

 

6.4 %

 

 

 

 

 

 

 

 

 

 

 

20 GHz to 26.5 GHz

 

5.5 %

 

 

 

 

 

 

 

 

 

 

 

26.5 GHz to 40.8 GHz

 

6.5 %

 

 

 

 

 

 

100

mW

to

2

W

 

 

 

 

 

 

 

 

 

(+20

dBm

to

+33

dBm)

100 kHz to 300 kHz

 

6.0 %

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

6.0 %

 

 

 

 

 

 

 

 

 

 

 

1 MHz to 10 MHz

 

5.2 %

 

 

 

 

 

 

 

 

 

 

 

10 MHz to 500 MHz

 

4.8 %

 

 

 

 

 

 

 

 

 

 

 

500 MHz to 1 GHz

 

5.9 %

 

 

 

 

 

 

 

 

 

 

 

1 GHz to 4.2 GHz

 

6.7 %

 

 

 

 

 

 

2
(+33

mW
dBm

to

10
+40

W
dBm)

100 kHz to 300 kHz

 

7.3 %

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

7.3 %

 

 

 

 

 

 

 

 

 

 

 

1 MHz to 10 MHz

 

6.6 %

 

 

 

 

 

 

 

 

 

 

 

10 MHz to 500 MHz

 

6.3 %

 

 

 

 

 

 

 

 

 

 

 

500 MHz to 1 GHz

 

7.1 %

 

 

 

 

 

 

 

 

 

 

 

1 GHz to 4.2 GHz

 

7.6 %

 

 

 

 

 

 

10
(+40

W
dBm

to

25
+44

W
dBm)

100 kHz to 300 kHz

 

7.4 %

 

 

 

 

 

 

 

 

 

 

 

300 kHz to 1 MHz

 

7.4 %

 

 

 

 

 

 

 

 

 

 

 

1 MHz to 10 MHz

 

6.7 %

 

 

 

 

 

 

 

 

 

 

 

10 MHz to 500 MHz

 

6.2 %

 

 

 

 

 

 

 

 

 

 

 

500 MHz to 1 GHz

 

6.9 %

 

 

 

 

 

 

 

 

 

 

 

1 GHz to 2 GHz

 

8.0 %

 

 

 

 

 

 

 

 

 

 

 

2 GHz to 4.2 GHz

 

9.5 %

 

 

 

 

19*

HF attenuation / attenuators, attenuation pads, selective HF meters, spectral and circuit analyzers, testers, radiocommunication and radio navigation testers, simulators, imitators (ATC, DME, TACAN, TCAS, VOR-ILS)

0

dBm

to

10

dBm

100 kHz to 400 kHz

 

 

0.048 dB

Indirect measurement power method

KP 16/E

1

 

 

 

 

 

 

 

 

400 kHz to 1.3 GHz

 

 

0.048 dB

 

 

 

 

 

 

 

 

 

 

 

1.3 GHz to 3 GHz

 

 

0.048 dB

 

 

 

 

 

 

 

 

 

 

 

3 GHz to 6 GHz

 

 

0.059 dB

 

 

 

 

 

 

10

dBm

to

20

dBm

100 kHz to 400 kHz

 

 

0.064 dB

 

 

 

 

 

 

 

 

 

 

 

400 kHz to 1.3 GHz

 

 

0.067 dB

 

 

 

 

 

 

 

 

 

 

 

1.3 GHz to 3 GHz

 

 

0.062 dB

 

 

 

 

 

 

 

 

 

 

 

3 GHz to 6 GHz

 

 

0.076 dB

 

 

 

 

 

 

20

dBm

to

30

dBm

100 kHz to 400 kHz

 

 

0.11 dB

 

 

 

 

 

 

 

 

 

 

 

400 kHz to 1.3 GHz

 

 

0.11 dB

 

 

 

 

 

 

 

 

 

 

 

1.3 GHz to 3 GHz

 

 

0.11 dB

 

 

 

 

 

 

 

 

 

 

 

3 GHz to 6 GHz

 

 

0.11 dB

 

 

 

 

 

 

30

dBm

to

40

dBm

100 kHz to 400 kHz

 

 

0.15 dB

 

 

 

 

 

 

 

 

 

 

 

400 kHz to 3 GHz

 

 

0.15 dB

 

 

 

 

 

 

 

 

 

 

 

3 GHz to 6 GHz

 

 

0.17 dB

 

 

 

 

 

 

40

dBm

to

60

dBm

100 kHz to 3 GHz

 

 

0.18 dB

 

 

 

 

 

 

 

 

 

 

 

3 GHz to 6 GHz

 

 

0.19 dB

 

 

 

 

 

 

60

dBm

to

65

dBm

100 kHz to 6 GHz

 

 

0.25 dB

 

 

 

 

 

19*

HF attenuation / attenuators, attenuation pads, selective HF meters, spectral and circuit analyzers, testers, radiocommunication and radio navigation testers, simulators, imitators (ATC, DME, TACAN, TCAS, VOR-ILS)

0

dBm

to

10

dBm

100 kHz to 400 kHz

 

 

0.048 dB

Indirect measurement power method

KP 16/E

1

 

 

 

 

 

 

 

400 kHz to 1.3 GHz

 

 

0.048 dB

 

 

 

 

 

 

 

 

 

 

1.3 GHz to 3 GHz

 

 

0.048 dB

 

 

 

 

 

 

 

 

 

 

3 GHz to 6 GHz

 

 

0.059 dB

 

 

 

 

 

10

dBm

to

20

dBm

100 kHz to 400 kHz

 

 

0.064 dB

 

 

 

 

 

 

 

 

 

 

400 kHz to 1.3 GHz

 

 

0.067 dB

 

 

 

 

 

 

 

 

 

 

1.3 GHz to 3 GHz

 

 

0.062 dB

 

 

 

 

 

 

 

 

 

 

3 GHz to 6 GHz

 

 

0.076 dB

 

 

 

 

 

20

dBm

to

30

dBm

100 kHz to 400 kHz

 

 

0.11 dB

 

 

 

 

 

 

 

 

 

 

400 kHz to 1.3 GHz

 

 

0.11 dB

 

 

 

 

 

 

 

 

 

 

1.3 GHz to 3 GHz

 

 

0.11 dB

 

 

 

 

 

 

 

 

 

 

3 GHz to 6 GHz

 

 

0.11 dB

 

 

 

 

 

30

dBm

to

40

dBm

100 kHz to 400 kHz

 

 

0.15 dB

 

 

 

 

 

 

 

 

 

 

400 kHz to 3 GHz

 

 

0.15 dB

 

 

 

 

 

 

 

 

 

 

3 GHz to 6 GHz

 

 

0.17 dB

 

 

 

 

 

40

dBm

to

60

dBm

100 kHz to 3 GHz

 

 

0.18 dB

 

 

 

 

 

 

 

 

 

 

3 GHz to 6 GHz

 

 

0.19 dB

 

 

 

 

 

60

dBm

to

65

dBm

100 kHz to 6 GHz

 

 

0.25 dB

 

 

 

20*

Modulation swing of frequency modulation / generators, modulation meters, selective HF meters, spectral analyzers, radiocommunication and radio navigation testers, simulators and imitators (ATC, DME, TACAN, TCAS, VOR-ILS)

 

 

 

 

 

Modulation frequency

Carrier frequency

 

 

Direct measurement or comparison with a modulation analyzer

KP 21/E

1

 

 

0.1

kHz

to

0.2

kHz

30 Hz to 60 kHz

(0.03 to 120) MHz

3.4 %

 

 

 

 

 

 

 

 

 

 

 

 

(120 to 550) MHz

3.4 %

 

 

 

 

 

 

 

 

 

 

 

 

(0.55 to 1) GHz

3.2 %

 

 

 

 

 

 

 

 

 

 

 

(60 to 100) kHz

(4.25 to 120) MHz

3.4 %

 

 

 

 

 

 

 

 

 

 

 

 

(120 to 550) MHz

6.8 %

 

 

 

 

 

 

 

 

 

 

 

 

(0.55 to 1) GHz

13 %

 

 

 

 

 

 

0.2

kHz

to

0.3

kHz

30 Hz to 60 kHz

(0.03 to 120) MHz

2.4 %

 

 

 

 

 

 

 

 

 

 

 

 

(120 to 550) MHz

2.4 %

 

 

 

 

 

 

 

 

 

 

 

 

(0.55 to 1) GHz

3.2 %

 

 

 

 

 

 

 

 

 

 

 

(60 to 100) kHz

(4.25 to 120) MHz

2.4 %

 

 

 

 

 

 

 

 

 

 

 

 

(120 to 550) MHz

4 %

 

 

 

 

 

 

 

 

 

 

 

 

(0.55 to 1) GHz

6.8 %

 

 

 

 

 

 

0.3

kHz

to

0.4

kHz

30 Hz to 60 kHz

(0.03 to 120) MHz

2 %

 

 

 

 

 

 

 

 

 

 

 

 

(120 to 550) MHz

2 %

 

 

 

 

 

 

 

 

 

 

 

 

(0.55 to 1) GHz

4.9 %

 

 

 

 

 

 

 

 

 

 

 

(60 to 100) kHz

(4.25 to 120) MHz

2 %

 

 

 

 

 

 

 

 

 

 

 

 

(120 to 550) MHz

3.1 %

 

 

 

 

 

 

 

 

 

 

 

 

(0.55 to 1) GHz

4.9 %

 

 

 

 

 

 

0.4

kHz

to

0.5

kHz

30 Hz to 60 kHz

(0.03 to 120) MHz

1.9 %