Scope of accreditation ACL 2222
RELATED LINKS
The current scope of accreditation of Accredited Calibration Laboratory No. 2222
(note: content is not optimized for mobile devices, we recommend switching to "desktop" view)
The Appendix is an integral part of Certificate of Accreditation No. 21/2022 of 07/01/2022
Accredited entity according to ČSN EN ISO/IEC 17025:2018
Institut pro testování a certifikaci, a. s., Calibration Laboratory, Sokolovská 573, 686 01 Uherské Hradiště
Calibration laboratory locations:
1. Uherské Hradiště Sokolovská 573, 686 01 Uherské Hradiště
2. Zlín třída Tomáše Bati 5264, 760 01 Zlín
3. Praha Křelovická 970, 104 00 Praha 10 – Uhříněves
4. Brno Křižíkova 70, 612 00 Brno
CMC for the field of measured quantity: Length
Ord. number1 |
Calibrated quantity / Subject of calibration |
Nominal range |
Parameter(s) of the meas. quantity |
Lowest expanded measurement uncertainty specified2 |
Calibration principle |
Calibration procedure identification3 |
Work-place |
||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
min. |
unit |
|
max. |
unit |
|||||||||||
1 |
Slide gauges |
0 |
mm |
to |
100 |
mm |
|
|
|
12 mm |
Direct measurement using parallel gauge blocks |
KP 1/D |
1 |
||
|
|
100 |
mm |
to |
500 |
mm |
|
|
|
14 mm |
|
|
|
||
2 |
Depth gauges |
0 |
mm |
to |
200 |
mm |
|
|
|
13 mm |
Direct measurement using parallel gauge blocks |
KP 1/D |
|
||
3 |
Height gauges |
0 |
mm |
to |
500 |
mm |
|
|
|
14 mm |
Direct measurement using parallel gauge blocks |
KP 1/D |
|
||
4 |
Micrometers |
0 |
mm |
to |
50 |
mm |
|
|
|
1.6 mm |
Direct measurement using parallel gauge blocks |
KP 2/D |
|
||
|
|
50 |
mm |
to |
100 |
mm |
|
|
|
2.0 mm |
|
|
|
||
|
|
100 |
mm |
to |
200 |
mm |
|
|
|
3.2 mm |
|
|
|
||
5 |
Micrometer depth gauges |
0 |
mm |
to |
100 |
mm |
|
|
|
2 mm |
Direct measurement using parallel gauge blocks |
KP 2/D |
|
||
6 |
Three-contact internal gauge with a micrometer screw |
0 |
mm |
to |
50 |
mm |
|
|
|
1.6 mm |
Direct measurement of standard measuring rings. |
KP 2/D |
|
||
|
|
50 |
mm |
to |
75 |
mm |
|
|
|
1.8 mm |
|
|
|
||
|
|
75 |
mm |
to |
100 |
mm |
|
|
|
2.0 mm |
|
|
|
||
7 |
Three-contact internal gauge with a lever mechanism |
0 |
mm |
to |
50 |
mm |
|
|
|
2.1 mm |
Direct measurement of standard measuring rings |
KP 2/D |
|
||
|
|
50 |
mm |
to |
75 |
mm |
|
|
|
2.3 mm |
|
|
|
||
|
|
75 |
mm |
to |
100 |
mm |
|
|
|
2.5 mm |
|
|
|
||
8 |
Two-contact internal gauge |
0 |
mm |
to |
35 |
mm |
|
|
|
2.5 mm |
Direct measurement of standard measuring rings |
KP 2/D |
|
||
|
|
35 |
mm |
to |
65 |
mm |
|
|
|
2.6 mm |
|
|
|
||
|
|
65 |
mm |
to |
75 |
mm |
|
|
|
2.7 mm |
|
|
|
||
|
|
75 |
mm |
to |
100 |
mm |
|
|
|
2.9 mm |
|
|
|
||
9 |
Pasameters |
0 |
mm |
to |
25 |
mm |
|
|
|
0.5 mm |
Direct measurement using parallel gauge blocks |
KP 2/D |
1 |
||
|
|
25 |
mm |
to |
50 |
mm |
|
|
|
0.8 mm |
|
|
|
||
|
|
50 |
mm |
to |
75 |
mm |
|
|
|
1.2 mm |
|
|
|
||
|
|
75 |
mm |
to |
100 |
mm |
|
|
|
1.5 mm |
|
|
|
||
|
|
100 |
mm |
to |
125 |
mm |
|
|
|
1.8 mm |
|
|
|
||
|
|
125 |
mm |
to |
150 |
mm |
|
|
|
2.2 mm |
|
|
|
||
|
|
150 |
mm |
to |
175 |
mm |
|
|
|
2.5 mm |
|
|
|
||
|
|
175 |
mm |
to |
200 |
mm |
|
|
|
2.9 mm |
|
|
|
||
10 |
Micropasameters |
0 |
mm |
to |
25 |
mm |
|
|
|
2.0 mm |
Direct measurement using parallel gauge blocks |
KP 2/D |
|
||
|
|
25 |
mm |
to |
50 |
mm |
|
|
|
2.5 mm |
|
|
|
||
|
|
50 |
mm |
to |
75 |
mm |
|
|
|
3.2 mm |
|
|
|
||
|
|
75 |
mm |
to |
100 |
mm |
|
|
|
3.8 mm |
|
|
|
||
|
|
100 |
mm |
to |
125 |
mm |
|
|
|
4.5 mm |
|
|
|
||
|
|
125 |
mm |
to |
150 |
mm |
|
|
|
5.4 mm |
|
|
|
||
|
|
150 |
mm |
to |
175 |
mm |
|
|
|
6.1 mm |
|
|
|
||
|
|
175 |
mm |
to |
200 |
mm |
|
|
|
6.9 mm |
|
|
|
||
11 |
Centesimal indicators |
0 |
mm |
to |
100 |
mm |
|
|
(2.L |
+6) mm |
Comparison with a standard length gauge |
KP 3/D |
|
||
12 |
Millesimal indicators |
0 |
mm |
to |
50 |
mm |
|
|
(2.L |
+1) mm |
Comparison with a standard length gauge |
KP 3/D |
|
||
13 |
Cylindrical gauges |
0 |
mm |
to |
200 |
mm |
|
|
(3.L |
+0.5) mm |
Direct measurement by a length gauge |
KP 3/D |
|
||
14 |
Slot gauges |
0 |
mm |
to |
200 |
mm |
|
|
(3.L |
+0.5) mm |
Direct measurement by a length gauge |
KP 3/D |
|
||
15 |
Limit plug gauges |
0 |
mm |
to |
200 |
mm |
|
|
(5.L |
+3) mm |
Direct measurement by a length gauge |
KP 3/D |
|
||
16 |
Precision plug gauges |
0 |
mm |
to |
50 |
mm |
|
|
(1.L |
+0.5) mm |
Direct measurement by a length gauge |
KP 3/D |
|
||
17 |
Measuring wires |
0.17 |
mm |
to |
6.35 |
mm |
|
|
(1.L |
+0.5) mm |
Direct measurement by a length gauge |
KP 3/D |
|
||
18 |
Setting rings |
10 |
mm |
to |
200 |
mm |
|
|
(4.L |
+1.2) mm |
Direct measurement by a length gauge |
KP 3/D |
|
||
19 |
Micrometer depth gauges |
0 |
mm |
to |
100 |
mm |
|
|
|
2.0 mm |
Direct measurement by a length gauge |
KP 3/D |
|
||
20 |
Steel rules |
0 |
mm |
to |
250 |
mm |
|
|
|
0.05 mm |
Direct measurement by a standard microscope |
KP 4/D |
|
||
21 |
Rules – straightness |
0 |
mm |
to |
250 |
mm |
|
|
|
0.007 mm |
Direct measurement by a standard microscope. |
KP 4/D |
|
||
22 |
Thread gauges |
0 |
mm |
to |
300 |
mm |
|
|
|
0.01 mm |
Direct measurement by a standard microscope. |
KP 4/D |
1 |
||
23 |
Radius gauges |
0 |
mm |
to |
300 |
mm |
|
|
|
0.01 mm |
Direct measurement by a standard microscope. |
KP 4/D |
|
||
24 |
Profile gauges |
0 |
mm |
to |
300 |
mm |
|
|
(1.L |
+0.5) mm |
Direct measurement by a standard microscope. |
KP 4/D |
|
||
25 |
Parallel gauge blocks |
0.5 |
mm |
to |
100 |
mm |
|
|
(2.L |
+0.2) mm |
Comparison with standard parallel gauge blocks |
KP 21/D |
|
||
1) Asterisk at the ordinal number identifies the calibrations, which the Laboratory is qualified to carry out outside the permanent laboratory premises.
2 The expanded measurement uncertainty is in accordance with ILAC-P14 and EA-4/02, part of CMC, and it is the lowest value of the respective uncertainty. If not stated otherwise, its coverage probability is approx. 95 %. If not stated otherwise, the uncertainty values stated without a unit are relative to the value measured. If the calibration is carried out outside the laboratory premises, the measurement uncertainty may be affected.
3 If the document identifying the calibration procedure is dated, only these specific procedures are used. If the document identifying the calibration procedure is not dated, the latest edition of the specified procedure is used (including any changes).
L measured length in metres
CMC for the field of measured quantity: Plane angle
Ord. number1 |
Calibrated quantity / Subject of calibration |
Nominal range |
Parameter(s) of the meas. quantity |
Lowest expanded measurement uncertainty specified2 |
Calibration principle |
Calibration procedure identification3 |
Work-place |
|||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
min. |
unit |
|
max. |
unit |
||||||||||
1 |
Angles |
0 |
° |
to |
180 |
° |
|
|
|
2.5 ¢ |
Direct measurement by a standard microscope. |
KP 4/D |
1 |
|
2 |
Angle gauges |
0 |
° |
to |
360 |
° |
|
|
|
2.5 ¢ |
Direct measurement by a standard microscope. |
KP 4/D |
|
|
1) Asterisk at the ordinal number identifies the calibrations, which the Laboratory is qualified to carry out outside the permanent laboratory premises.
2 The expanded measurement uncertainty is in accordance with ILAC-P14 and EA-4/02, part of CMC, and it is the lowest value of the respective uncertainty. If not stated otherwise, its coverage probability is approx. 95 %. If not stated otherwise, the uncertainty values stated without a unit are relative to the value measured. If the calibration is carried out outside the laboratory premises, the measurement uncertainty may be affected.
3 If the document identifying the calibration procedure is dated, only these specific procedures are used. If the document identifying the calibration procedure is not dated, the latest edition of the specified procedure is used (including any changes).
CMC for the field of measured quantity: Mechanical motion (speed)
Ord. number1 |
Calibrated quantity / Subject of calibration |
Nominal range |
Parameter(s) of the meas. quantity |
Lowest expanded measurement uncertainty specified2 |
Calibration principle |
Calibration procedure identification3 |
Work-place |
||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
min. |
unit |
|
max. |
unit |
|||||||||
1. |
Speed / analogue and digital contact and contactless, revolution counters, rpm sensors, stroboscopes. |
1 |
min-1 |
to |
10,000 |
min-1 |
|
|
|
0.006 min-1 |
Direct measurement of optical signal or direct comparison with a standard revolution counter |
KP 6/F |
1 |
|
|
10,000 |
min-1 |
to |
999,999 |
min-1 |
|
|
|
0.03 min-1 |
|
|
|
|
|
2 |
min-1 |
to |
999.99 |
min-1 |
|
|
|
0.1 min-1 |
Measurement by a standard revolution counter |
|
|
|
|
1,000 |
min-1 |
to |
9,999 |
min-1 |
|
|
|
0.9 min-1 |
|
|
|
|
|
10,000 |
min-1 |
to |
999,999 |
min-1 |
|
|
|
7.0 min-1 |
|
|
|
1) Asterisk at the ordinal number identifies the calibrations, which the Laboratory is qualified to carry out outside the permanent laboratory premises.
2 The expanded measurement uncertainty is in accordance with ILAC-P14 and EA-4/02, part of CMC, and it is the lowest value of the respective uncertainty. If not stated otherwise, its coverage probability is approx. 95 %. If not stated otherwise, the uncertainty values stated without a unit are relative to the value measured. If the calibration is carried out outside the laboratory premises, the measurement uncertainty may be affected.
3 If the document identifying the calibration procedure is dated, only these specific procedures are used. If the document identifying the calibration procedure is not dated, the latest edition of the specified procedure is used (including any changes).
CMC for the field of measured quantity: Hardness
Ord. number1 |
Calibrated quantity / Subject of calibration |
Nominal range |
Parameter(s) of the meas. quantity |
Lowest expanded measurement uncertainty specified2 |
Calibration principle |
Calibration procedure identification3 |
Work-place |
||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
min. |
unit |
|
max. |
unit |
|||||||||||
1 |
Hardness tester Shore A |
1 |
Shore A |
to |
100 |
Shore A |
Spring force 31 mN |
|
|
0.3 Shore A |
Direct measurement of force on a calibrator and measurement of tip geometry using an image analyzer |
KP 1/H |
2 |
||
2 |
Hardness tester Shore C |
1 |
Shore C |
to |
100 |
Shore C |
Spring force 71 mN |
|
|
0.3 Shore C |
|
|
|
||
3 |
Hardness tester Shore D |
1 |
Shore D |
to |
100 |
Shore D |
Spring force 71 mN |
|
|
0.3 Shore D |
|
|
|
||
4 |
Hardness tester IRHD/N |
10 |
IRHD |
to |
100 |
IRHD |
|
|
|
0.6 IRHD |
Direct measurement of force on a calibrator and measurement of tip geometry using an image analyzer |
KP 2/H |
|
||
5 |
Hardness tester IRHD/m |
10 |
IRHD |
to |
100 |
IRHD |
|
|
|
1.1 IRHD |
|
|
|
||
1 Asterisk at the ordinal number identifies the calibrations, which the Laboratory is qualified to carry out outside the permanent laboratory premises.
2 The expanded measurement uncertainty is in accordance with ILAC-P14 and EA-4/02, part of CMC, and it is the lowest value of the respective uncertainty. If not stated otherwise, its coverage probability is approx. 95 %. If not stated otherwise, the uncertainty values stated without a unit are relative to the value measured. If the calibration is carried out outside the laboratory premises, the measurement uncertainty may be affected.
3 If the document identifying the calibration procedure is dated, only these specific procedures are used. If the document identifying the calibration procedure is not dated, the latest edition of the specified procedure is used (including any changes).
CMC for the field of measured quantity: Pressure, mechanical stress
Ord. number1 |
Calibrated quantity / Subject of calibration |
Nominal range |
Parameter(s) of the meas. quantity |
Lowest expanded measurement uncertainty specified2 |
Calibration principle |
Calibration procedure identification3 |
Work-place |
||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
min. |
unit |
|
max. |
unit |
|||||||||
1. |
Deformation pressure gauges |
0 |
kPa |
to |
10 |
kPa |
Overpressure |
gas |
0.015 % |
+ 0.058 kPa |
Comparison with a standard digital manometer |
KP 1/P |
1, 4 |
|
|
10 |
kPa |
to |
400 |
kPa |
|
|
0.046 % |
+ 0.07 kPa |
|
|
|
|
|
0.4 |
MPa |
to |
2 |
MPa |
|
|
0.046 % |
+ 0.18 kPa |
|
|
|
|
|
2 |
MPa |
to |
6 |
MPa |
|
|
0.015 % |
+ 0.42 kPa |
|
|
|
|
|
0 |
kPa |
to |
95 |
kPa |
Underpressure |
gas |
0.046 % |
+ 0.07 kPa |
|
|
|
|
|
70 |
kPa |
to |
130 |
kPa |
Absolute pressure |
gas4 |
0.015 % |
+ 0.12 kPa |
|
|
|
|
|
130 |
kPa |
to |
520 |
kPa |
|
|
0.046 % |
+ 0.13 kPa |
|
|
|
|
|
0.52 |
MPa |
to |
2.12 |
MPa |
|
|
0.046 % |
+ 0.21 kPa |
|
|
|
|
|
2.12 |
MPa |
to |
6 |
MPa |
|
|
0.015 % |
+ 0.44 kPa |
|
|
|
|
|
6 |
MPa |
to |
16 |
MPa |
Overpressure, absolute pressure |
gas, liquid4 |
0.015 % |
+ 1.2 kPa |
|
|
|
|
|
16 |
MPa |
to |
60 |
MPa |
Overpressure, absolute pressure |
liquid4 |
0.046 % |
+ 11 kPa |
|
|
|
|
|
60 |
MPa |
to |
100 |
MPa |
|
|
0.012 % |
+ 8.3 kPa |
|
|
|
|
|
70 |
kPa |
to |
120 |
kPa |
Barometric pressure |
|
|
0.082 kPa |
|
|
|
2* |
Pressure converters |
0 |
kPa |
to |
10 |
kPa |
Overpressure |
gas |
0.015 % |
+ 0.0018 kPa |
Comparison with a standard digital manometer |
KP 2/P |
|
|
|
10 |
kPa |
to |
400 |
kPa |
|
|
0.048 % |
+ 0.035 kPa |
|
|
|
|
|
0.4 |
MPa |
to |
2 |
MPa |
|
|
0.048 % |
+ 0.12 kPa |
|
|
|
|
|
2 |
MPa |
to |
6 |
MPa |
|
|
0.015 % |
+ 0.35 kPa |
|
|
|
|
|
0 |
kPa |
to |
95 |
kPa |
Underpressure |
gas |
0.048 % |
+ 0.035 kPa |
|
|
|
|
|
70 |
kPa |
to |
130 |
kPa |
Absolute pressure |
gas4 |
0.015 % |
+ 0.1 kPa |
|
|
|
|
|
130 |
kPa |
to |
520 |
kPa |
|
|
0.048 % |
+ 0.11 kPa |
|
|
|
|
|
0.52 |
MPa |
to |
2.12 |
MPa |
|
|
0.048 % |
+ 0.16 kPa |
|
|
|
|
|
2.12 |
MPa |
to |
6 |
MPa |
|
|
0.015 % |
+ 0.36 kPa |
|
|
|
|
|
6 |
MPa |
to |
16 |
MPa |
Overpressure, absolute pressure |
gas, liquid4 |
0.015 % |
+ 0.93 kPa |
|
|
|
|
|
16 |
MPa |
to |
60 |
MPa |
Overpressure, absolute pressure |
liquid4 |
0.048 % |
+ 11 kPa |
|
|
|
|
|
60 |
MPa |
to |
100 |
MPa |
|
|
0.012 % |
+ 8.1 kPa |
|
|
|
|
|
70 |
kPa |
to |
120 |
kPa |
Barometric pressure |
|
0.12 % |
|
|
|
|
3* |
Digital manometers |
0 |
kPa |
to |
10 |
kPa |
Overpressure |
gas |
0.015 % |
+ 0.0018 kPa |
Comparison with a standard digital manometer |
KP 3/P |
|
|
|
10 |
kPa |
to |
400 |
kPa |
|
|
0.046 % |
+ 0.039 kPa |
|
|
|
|
|
0.4 |
MPa |
to |
2 |
MPa |
|
|
0.046 % |
+ 0.14 kPa |
|
|
|
|
|
2 |
MPa |
to |
6 |
MPa |
|
|
0.015 % |
+ 0.36 kPa |
|
|
|
|
|
0 |
kPa |
to |
95 |
kPa |
Underpressure |
gas |
0.046 % |
+ 0.039 kPa |
|
|
|
|
|
70 |
kPa |
to |
130 |
kPa |
Absolute pressure |
gas4 |
0.015 % |
+ 0.1 kPa |
|
|
|
|
|
130 |
kPa |
to |
520 |
kPa |
|
|
0.046 % |
+ 0.11 kPa |
|
|
|
|
|
0.52 |
MPa |
to |
2.12 |
MPa |
|
|
0.046 % |
+ 0.17 kPa |
|
|
|
|
|
2.12 |
MPa |
to |
6 |
MPa |
|
|
0.015 % |
+ 0.37 kPa |
|
|
|
|
|
6 |
MPa |
to |
16 |
MPa |
Overpressure, absolute pressure |
gas, liquid4 |
0.015 % |
+ 0.98 kPa |
|
|
|
|
|
16 |
MPa |
to |
60 |
MPa |
Overpressure, absolute pressure |
liquid4 |
0.046 % |
+ 0.11 kPa |
|
|
|
|
|
60 |
MPa |
to |
100 |
MPa |
|
|
0.012 % |
+ 8.2 kPa |
|
|
|
|
|
70 |
kPa |
to |
120 |
kPa |
Barometric pressure |
|
|
0.058 kPa |
|
|
|
1) Asterisk at the ordinal number identifies the calibrations, which the Laboratory is qualified to carry out outside the permanent laboratory premises.
2 The expanded measurement uncertainty is in accordance with ILAC-P14 and EA-4/02, part of CMC, and it is the lowest value of the respective uncertainty. If not stated otherwise, its coverage probability is approx. 95 %. If not stated otherwise, the uncertainty values stated without a unit are relative to the value measured. If the calibration is carried out outside the laboratory premises, the measurement uncertainty may be affected.
3 If the document identifying the calibration procedure is dated, only these specific procedures are used. If the document identifying the calibration procedure is not dated, the latest edition of the specified procedure is used (including any changes).
4 The resulting pressure is the sum of relative and barometric pressure. The resulting uncertainty is the combination of relative and barometric pressure uncertainty.
CMC for the field of measured quantity: Temperature
Ord. num-ber1 |
Calibrated quantity / Subject of calibration |
Nominal range |
Parameter(s) of the meas. quantity |
Lowest expanded measurement uncertainty specified2 |
Calibration principle |
Calibration procedure identification3 |
Work-place |
||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
min. |
unit |
|
max. |
unit |
|||||||||
1* |
Thermocouples |
-40 |
°C |
to |
0 |
°C |
|
|
|
0.6 °C |
Comparison with a standard resistance sensor in liquid bath. |
KP 1/T |
1, 4 |
|
|
|
|
|
0 |
°C |
|
|
|
0.6 °C |
Comparison with a standard resistance sensor in dry block. |
|
|
|
|
0 |
°C |
to |
100 |
°C |
|
|
|
0.6 °C |
|
|
|
|
|
100 |
°C |
to |
250 |
°C |
|
|
|
0.7 °C |
|
|
|
|
|
250 |
°C |
to |
600 |
°C |
|
|
|
0.8 °C |
|
|
|
|
|
600 |
°C |
to |
800 |
°C |
|
|
|
1.5 °C |
Comparison with a standard thermoelectric sensor in a horizontal oven |
|
|
|
|
800 |
°C |
to |
1,200 |
°C |
|
|
|
2.2 °C |
|
|
|
2* |
Resistance temperature sensors, resistance temperature sensors with a transducer |
-80 |
°C |
to |
-40 |
°C |
|
|
|
0.20 °C |
Comparison with a standard thermoelectric sensor in dry block |
KP 2/T |
4 |
|
|
-40 |
°C |
to |
0 |
°C |
|
|
|
0.07 °C |
Comparison with a standard resistance sensor in liquid bath. |
|
|
|
|
|
|
|
0 |
°C |
|
|
|
0.05 °C |
|
|
|
|
|
0 |
°C |
to |
100 |
°C |
|
|
|
0.08 °C |
|
|
|
|
|
100 |
°C |
to |
250 |
°C |
|
|
|
0.10 °C |
|
|
|
|
|
250 |
°C |
to |
350 |
°C |
|
|
|
0.43 °C |
Comparison with a standard resistance sensor in dry block. |
|
|
|
|
350 |
°C |
to |
600 |
°C |
|
|
|
0.62 °C |
|
|
|
3* |
Glass thermometers |
-40 |
°C |
to |
0 |
°C |
|
|
|
0.20 °C |
Comparison with a standard resistance sensor in liquid bath. |
KP 3/T |
1, 4 |
|
|
|
|
|
0 |
°C |
|
|
|
0.05 °C |
|
|
|
|
|
0 |
°C |
to |
100 |
°C |
|
|
|
0.14 °C |
|
|
|
|
|
100 |
°C |
to |
250 |
°C |
|
|
|
0.15 °C |
|
|
|
|
|
250 |
°C |
to |
300 |
°C |
|
|
|
0.30 °C |
|
|
|
4* |
Direct-indicating thermometers |
-40 |
°C |
to |
0 |
°C |
|
|
|
0.07 °C |
Comparison with a standard resistance sensor in liquid bath. |
KP 4/T |
1, 4 |
|
|
|
|
|
0 |
°C |
|
|
|
0.06 °C |
|
|
|
|
|
0 |
°C |
to |
100 |
°C |
|
|
|
0.09 °C |
|
|
|
|
|
100 |
°C |
to |
250 |
°C |
|
|
|
0.10 °C |
|
|
|
|
|
250 |
°C |
to |
350 |
°C |
|
|
|
0.44 °C |
Comparison with a standard resistance sensor in dry block. |
|
|
|
|
350 |
°C |
to |
600 |
°C |
|
|
|
0.62 °C |
|
|
|
|
|
600 |
°C |
to |
800 |
°C |
|
|
|
1.6 °C |
Comparison with a standard thermoelectric sensor in a horizontal oven |
|
|
|
|
800 |
°C |
to |
1,200 |
°C |
|
|
|
2.2 °C |
|
|
|
5* |
Controllers and indicators |
0 |
°C |
to |
1,370 |
°C |
Type K |
|
|
0.3 °C |
Simulation of thermoelectric temperature sensors using DC voltage |
KP 5/T |
1, 4 |
|
|
0 |
°C |
to |
1,780 |
°C |
Type S |
|
|
0.6 °C |
|
|
|
|
|
200 |
°C |
to |
500 |
°C |
Type B |
|
|
1.5 °C |
|
|
|
|
|
500 |
°C |
to |
1,820 |
°C |
Type B |
|
|
0.6 °C |
|
|
|
|
|
0 |
°C |
to |
1,768 |
°C |
Type R |
|
|
0.6 °C |
|
|
|
|
|
0 |
°C |
to |
1,200 |
°C |
Type J |
|
|
0.2 °C |
|
|
|
|
|
0 |
°C |
to |
1,300 |
°C |
Type N |
|
|
0.2 °C |
|
|
|
|
|
-50 |
°C |
to |
850 |
°C |
RTD |
|
|
0.05 °C |
Simulation of resistance temperature sensors using DC electrical resistors |
|
|
6* |
Non-contact thermometers |
-30 |
°C |
to |
23 |
°C |
|
|
|
2.2 °C |
Direct measurement on a black body, comparison with a standard contactless thermometer. |
KP 6/T |
1, 4 |
|
|
23 |
°C |
to |
150 |
°C |
|
|
|
2.1 °C |
|
|
|
|
|
150 |
°C |
to |
250 |
°C |
|
|
|
2.2 °C |
|
|
|
|
|
250 |
°C |
to |
350 |
°C |
|
|
|
2.5 °C |
|
|
|
|
|
350 |
°C |
to |
500 |
°C |
|
|
|
3.0 °C |
|
|
|
7* |
Direct-indicating thermometers integrated in conditioning and thermal chambers and baths |
-80 |
°C |
to |
250 |
°C |
|
|
|
1.0 °C |
Comparison with a standard digital thermometer in a chamber (characterisation) |
KP 22/T |
1, 4 |
|
|
250 |
°C |
to |
600 |
°C |
|
|
|
1.1 °C |
|
|
|
|
|
600 |
°C |
to |
700 |
°C |
|
|
|
1.7 °C |
|
|
|
|
|
700 |
°C |
to |
850 |
°C |
|
|
|
1.9 °C |
|
|
|
|
|
850 |
°C |
to |
1,200 |
°C |
|
|
|
2.9 °C |
|
|
|
8* |
Direct-indicating thermometers and temperature measuring chains integrated in conditioning and thermal chambers and baths |
-20 |
°C |
to |
250 |
°C |
|
|
|
1.0 °C |
Comparison with a standard digital thermometer in a chamber (characterisation according to the standard AMS 2750) |
KP 21/T |
1 |
|
|
250 |
°C |
to |
600 |
°C |
|
|
|
1.1 °C |
|
|
|
|
|
600 |
°C |
to |
700 |
°C |
|
|
|
1.7 °C |
|
|
|
|
|
700 |
°C |
to |
850 |
°C |
|
|
|
1.9 °C |
|
|
|
|
|
850 |
°C |
to |
1,000 |
°C |
|
|
|
2.9 °C |
|
|
|
9* |
Temperature controllers and indicators of conditioning and thermal chambers and baths |
-20 |
°C |
to |
1,370 |
°C |
Type K |
|
|
0.3 °C |
Simulation of thermoelectric temperature sensors using DC voltage (characterisation according to the standard AMS 2750) |
KP 21/T |
1 |
|
|
-20 |
°C |
to |
1,300 |
°C |
Type N |
|
|
0.2 °C |
|
|
|
|
|
0 |
°C |
to |
1,780 |
°C |
Type S |
|
|
0.6 °C |
|
|
|
|
|
200 |
°C |
to |
500 |
°C |
Type B |
|
|
1.5 °C |
|
|
|
|
|
500 |
°C |
to |
1,800 |
°C |
Type B |
|
|
0.6 °C |
|
|
|
1) Asterisk at the ordinal number identifies the calibrations, which the Laboratory is qualified to carry out outside the permanent laboratory premises.
2 The expanded measurement uncertainty is in accordance with ILAC-P14 and EA-4/02, part of CMC, and it is the lowest value of the respective uncertainty. If not stated otherwise, its coverage probability is approx. 95 %. If not stated otherwise, the uncertainty values stated without a unit are relative to the value measured. If the calibration is carried out outside the laboratory premises, the measurement uncertainty may be affected.
3 If the document identifying the calibration procedure is dated, only these specific procedures are used. If the document identifying the calibration procedure is not dated, the latest edition of the specified procedure is used (including any changes).
CMC for the field of measured quantity: Electrical quantities
Ord. num-ber1 |
Calibrated quantity / Subject of calibration |
Nominal range |
Parameter(s) of the meas. quantity |
Lowest expanded measurement uncertainty specified2 |
Calibration principle |
Calibration procedure identification3 |
Work-place |
|
|||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
min. |
unit |
|
max. |
unit |
|
||||||||||||||||||||||
1 |
Direct-current voltage/ direct-current voltage meters (digital and analogue multimeters, voltmeters, oscilloscopes, testing and inspection instruments and their parts) |
0 |
mV |
to |
1 |
mV |
|
|
|
1.3 µV |
Voltage generation by a standard calibrator |
KP 1/E, KP 2/E, KP 3/E, KP 18/E |
1 |
|
|||||||||||||
|
|
1 |
mV |
to |
20 |
mV |
|
|
|
1.7 µV |
|
|
|
|
|||||||||||||
|
|
20 |
mV |
to |
100 |
mV |
|
|
|
2.1 µV |
|
|
|
|
|||||||||||||
|
|
100 |
mV |
to |
200 |
mV |
|
|
|
3.4 µV |
|
|
|
|
|||||||||||||
|
|
200 |
mV |
to |
500 |
mV |
|
|
|
0.5 µV |
|
|
|
|
|||||||||||||
|
|
500 |
mV |
to |
900 |
mV |
|
|
0.0011 % |
|
|
|
|
|
|||||||||||||
|
|
0.9 |
V |
to |
1.5 |
V |
|
|
0.0009 % |
|
|
|
|
|
|||||||||||||
|
|
1.5 |
V |
to |
2 |
V |
|
|
0.0008 % |
|
|
|
|
|
|||||||||||||
|
|
2 |
V |
to |
4 |
V |
|
|
0.0007 % |
|
|
|
|
|
|||||||||||||
|
|
4 |
V |
to |
10 |
V |
|
|
0.00055 % |
|
|
|
|
|
|||||||||||||
|
|
10 |
V |
to |
20 |
V |
|
|
0.00051 % |
|
|
|
|
|
|||||||||||||
|
|
20 |
V |
to |
40 |
V |
|
|
0.0011 % |
|
|
|
|
|
|||||||||||||
|
|
40 |
V |
to |
100 |
V |
|
|
0.00094 % |
|
|
|
|
|
|||||||||||||
|
|
100 |
V |
to |
200 |
V |
|
|
0.00082 % |
|
|
|
|
|
|||||||||||||
|
|
200 |
V |
to |
400 |
V |
|
|
0.0014 % |
|
|
|
|
|
|||||||||||||
|
|
400 |
V |
to |
1,100 |
V |
|
|
0.0013 % |
|
|
|
|
|
|||||||||||||
|
|
1,100 |
V |
to |
6,000 |
V |
|
|
0.6 % |
|
|
|
|
|
|||||||||||||
2 |
Direct-current voltage / direct-current voltage sources and calibrators, multifunction calibrators, testing and inspection instruments and their parts, high voltage testers. |
0 |
mV |
to |
120 |
mV |
|
|
0.00096 % |
+ 0.71 mV |
Direct measurement by a standard multimeter |
KP 8/E |
1 |
|
|||||||||||||
|
|
120 |
mV |
to |
1,200 |
mV |
|
|
0.00072 % |
+ 4.6 mV |
|
|
|
|
|||||||||||||
|
|
1.2 |
V |
to |
12 |
V |
|
|
0.00060 % |
+ 4.6 mV |
|
|
|
|
|||||||||||||
|
|
12 |
V |
to |
120 |
V |
|
|
0.00088 % |
+ 29 mV |
|
|
|
|
|||||||||||||
|
|
120 |
V |
to |
1,100 |
V |
|
|
0.00098 % |
+ 82 mV |
|
|
|
|
|||||||||||||
|
|
1,100 |
V |
to |
6,000 |
V |
|
|
0.6 % |
|
Measurement by a standard multimeter with a HV probe |
|
|
|
|||||||||||||
3 |
Direct current / direct-current meters (digital and analogue multimeters, ammeters, testing and inspection instruments and their parts, current probes). |
0 |
mA |
to |
2 |
mA |
|
|
|
3.8 nA |
Current generation by a standard calibrator |
KP 1/E, KP 2/E, KP 3/E |
1 |
|
|||||||||||||
|
|
2 |
mA |
to |
10 |
mA |
|
|
|
4.5 nA |
|
|
|
|
|||||||||||||
|
|
10 |
mA |
to |
20 |
mA |
|
|
|
5.5 nA |
|
|
|
|
|||||||||||||
|
|
20 |
mA |
to |
40 |
mA |
|
|
0.027 % |
|
|
|
|
|
|||||||||||||
|
|
40 |
mA |
to |
100 |
mA |
|
|
0.019 % |
|
|
|
|
|
|||||||||||||
|
|
100 |
mA |
to |
200 |
mA |
|
|
0.015 % |
|
|
|
|
|
|||||||||||||
|
|
200 |
mA |
to |
200 |
mA |
|
|
0.011 % |
|
|
|
|
|
|||||||||||||
|
|
200 |
mA |
to |
400 |
mA |
|
|
0.027 % |
|
|
|
|
|
|||||||||||||
|
|
400 |
mA |
to |
1 |
A |
|
|
0.019 % |
|
|
|
|
|
|||||||||||||
|
|
1 |
A |
to |
2 |
A |
|
|
0.015 % |
|
|
|
|
|
|||||||||||||
|
|
2 |
A |
to |
3 |
A |
|
|
0.045 % |
|
|
|
|
|
|||||||||||||
|
|
3 |
A |
to |
4 |
A |
|
|
0.038 % |
|
Comparison with current measured by a standard shunt with multimeter |
|
|
|
|||||||||||||
|
|
4 |
A |
to |
30 |
A |
|
|
0.035 % |
|
|
|
|
|
|||||||||||||
|
|
30 |
A |
to |
50 |
A |
|
|
0.15 % |
|
|
|
|
|
|||||||||||||
|
|
50 |
A |
to |
100 |
A |
|
|
0.28 % |
|
|
|
|
|
|||||||||||||
|
Direct current/ clamp ammeters and multimeters, clamp current probes. |
1 |
A |
to |
5 |
A |
|
|
0.2 % |
|
Current generation by a standard calibrator |
|
|
|
|||||||||||||
|
|
5 |
A |
to |
750 |
A |
|
|
0.13 % |
|
Simulation of reference current using calibrator and current coil |
|
|
|
|||||||||||||
4 |
Direct current / current sources, testing and inspection instruments and their parts. |
0 |
mA |
to |
12 |
mA |
|
|
0.0045 % |
+ 6.0 nA |
Direct measurement by a standard multimeter |
KP 1/E, KP 2/E, KP 3/E |
1 |
|
|||||||||||||
|
|
12 |
mA |
to |
120 |
mA |
|
|
0.0050 % |
+ 6.7 nA |
|
|
|
|
|||||||||||||
|
|
120 |
mA |
to |
1.2 |
mA |
|
|
0.0050 % |
+ 4.5 nA |
|
|
|
|
|||||||||||||
|
|
1.2 |
mA |
to |
12 |
mA |
|
|
0.0044 % |
+ 0.045 mA |
|
|
|
|
|||||||||||||
|
|
12 |
mA |
to |
120 |
mA |
|
|
0.0065 % |
+ 0.45 mA |
|
|
|
|
|||||||||||||
|
|
120 |
mA |
to |
1.05 |
A |
|
|
0.016 % |
+ 9.8 mA |
|
|
|
|
|||||||||||||
|
|
1.05 |
A |
to |
12 |
A |
|
|
0.0073 % |
+ 25 mA |
Indirect measurement using a standard shunt and multimeter |
KP 8/E |
|
|
|||||||||||||
|
|
12 |
A |
to |
20 |
A |
|
|
0.0058 % |
+ 4 mA |
|
|
|
|
|||||||||||||
|
|
20 |
A |
to |
50 |
A |
|
|
0.042 % |
+ 61 mA |
|
|
|
|
|||||||||||||
|
|
50 |
A |
to |
100 |
A |
|
|
0.047 % |
|
|
|
|
|
|||||||||||||
|
Direct current / current sources |
1 |
A |
to |
40 |
A |
|
|
3.8 % |
|
Direct measurement by a standard clamp multimeter |
KP 8/E |
1 |
|
|||||||||||||
|
|
40 |
A |
to |
100 |
A |
|
|
5 % |
|
|
|
|
|
|||||||||||||
|
|
100 |
A |
to |
250 |
A |
|
|
4 % |
|
|
|
|
|
|||||||||||||
|
|
250 |
A |
to |
400 |
A |
|
|
2.5 % |
|
|
|
|
|
|||||||||||||
|
|
400 |
A |
to |
1,000 |
A |
|
|
2 % |
|
|
|
|
|
|||||||||||||
5 |
Alternating-current voltage/ analogue multimeters, voltmeters, oscilloscopes, testing and inspection instruments and their parts |
0.1 |
mV |
to |
1.0 |
mV |
10 Hz to 100 kHz |
|
|
8 mV |
Voltage generation by a standard calibrator |
KP 1/E, KP 2/E, KP 3/E, KP 18/E |
1 |
|
|||||||||||||
|
|
|
|
|
|
|
100 kHz to 300 kHz |
|
|
18 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
|
34 mV |
|
|
|
|
|||||||||||||
|
|
1.0 |
mV |
to |
10.0 |
mV |
10 Hz to 100 kHz |
|
|
15 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
100 kHz to 300 kHz |
|
|
30 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
|
70 mV |
|
|
|
|
|||||||||||||
|
|
10.0 |
mV |
to |
20.0 |
mV |
10 Hz to 30 kHz |
|
|
13mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
30 kHz to 100 kHz |
|
|
18 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
100 kHz to 300 kHz |
|
|
45 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
|
110 mV |
|
|
|
|
|||||||||||||
|
|
20.0 |
mV |
to |
40.0 |
mV |
10 Hz to 30 kHz |
|
|
17 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
30 kHz to 100 kHz |
|
|
29 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
100 kHz to 300 kHz |
|
|
75 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
|
270 mV |
|
|
|
|
|||||||||||||
|
|
40.0 |
mV |
to |
200.0 |
mV |
10 Hz to 30 kHz |
|
0.045 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
30 kHz to 100 kHz |
|
0.075 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
100 kHz to 300 kHz |
|
0.2 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
0.7 % |
|
|
|
|
|
|||||||||||||
|
|
200.0 |
mV |
to |
400.0 |
mV |
10 Hz to 300 Hz |
|
0.04 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 Hz to 30 kHz |
|
0.03 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
30 kHz to 100 kHz |
|
0.04 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
100 kHz to 300 kHz |
|
0.1 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
0.55 % |
|
|
|
|
|
|||||||||||||
|
|
0.4 |
V |
to |
1 |
V |
10 Hz to 300 Hz |
|
0.022 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 Hz to 30 kHz |
|
0.016 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
30 kHz to 100 kHz |
|
0.025 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
100 kHz to 300 kHz |
|
0.07 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
0.47 % |
|
|
|
|
|
|||||||||||||
|
|
1 |
V |
to |
2 |
V |
10 Hz to 300 Hz |
|
0.015 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 Hz to 30 kHz |
|
0.01 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
30 kHz to 100 kHz |
|
0.02 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
100 kHz to 300 kHz |
|
0.05 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
0.43 % |
|
|
|
|
|
|||||||||||||
|
|
2 |
V |
to |
4 |
V |
10 Hz to 300 Hz |
|
0.03 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 Hz to 30 kHz |
|
0.02 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
30 kHz to 100 kHz |
|
0.03 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
100 kHz to 300 kHz |
|
0.11 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
0.55 % |
|
|
|
|
|
|||||||||||||
|
|
4 |
V |
to |
10 |
V |
10 Hz to 300 Hz |
|
0.02 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 Hz to 30 kHz |
|
0.013 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
30 kHz to 100 kHz |
|
0.02 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
100 kHz to 300 kHz |
|
0.086 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
0.47 % |
|
|
|
|
|
|||||||||||||
|
|
10 |
V |
to |
20 |
V |
10 Hz to 300 Hz |
|
0.015 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 Hz to 30 kHz |
|
0.01 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
30 kHz to 100 kHz |
|
0.015 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
100 kHz to 300 kHz |
|
0.075 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
0.42 % |
|
|
|
|
|
|||||||||||||
|
|
20 |
V |
to |
40 |
V |
10 Hz to 300 Hz |
|
0.03 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 Hz to 30 kHz |
|
0.019 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
30 kHz to 100 kHz |
|
0.035 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
100 kHz to 300 kHz |
|
0.11 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
0.15 % |
|
|
|
|
|
|||||||||||||
|
|
40 |
V |
to |
100 |
V |
10 Hz to 300 Hz |
|
0.022 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 Hz to 10 kHz |
|
0.01 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
10 kHz to 30 kHz |
|
0.014 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
30 kHz to 100 kHz |
|
0.03 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
100 kHz to 200 kHz |
|
0.08 % |
|
|
|
|
|
|||||||||||||
|
|
100 |
V |
to |
200 |
V |
10 Hz to 300 Hz |
|
0.02 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 Hz to 30 kHz |
|
0.01 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
30 kHz to 100 kHz |
|
0.023 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
100 kHz to 200 kHz |
|
0.065 % |
|
|
|
|
|
|||||||||||||
|
|
200 |
V |
to |
750 |
V |
45 Hz to 300 Hz |
|
0.03 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 Hz to 10 kHz |
|
0.025 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
10 kHz to 30 kHz |
|
0.03 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
30 kHz to 100 kHz |
|
0.15 % |
|
|
|
|
|
|||||||||||||
|
|
750 |
V |
to |
1,100 |
V |
45 Hz to 300 Hz |
|
0.027 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 Hz to 10 kHz |
|
0.022 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
10 kHz to 30 kHz |
|
0.025 % |
|
|
|
|
|
|||||||||||||
|
|
1,100 |
V |
to |
5,000 |
V |
40 Hz to 100 Hz |
|
2 % |
|
Comparison with a standard multimeter with a HV probe |
|
|
|
|||||||||||||
6 |
Alternating-current voltage / voltage sources and calibrators, multifunction calibrators, testing and inspection instruments and their parts, high voltage testers |
0.1 |
mV |
to |
12 |
mV |
1 Hz to 40 Hz |
|
0.062 % |
+ 3,5 mV |
Direct measurement by a standard multimeter |
KP 8/E |
1 |
|
|||||||||||||
|
|
|
|
|
|
|
40 Hz to 1,000 Hz |
|
0.048 % |
+ 1,4 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 20 kHz |
|
0.048 % |
+ 1,4 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
20 kHz to 50 kHz |
|
0.13 % |
+ 1,4 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
50 kHz to 100 kHz |
|
0.6 % |
+ 1,4 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
100 kHz to 300 kHz |
|
4.7 % |
+ 2,4 mV |
|
|
|
|
|||||||||||||
|
|
12 |
mV |
to |
120 |
mV |
1 Hz to 40 Hz |
|
0.036 % |
+ 2,8 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
40 Hz to 1,000 Hz |
|
0.017 % |
+ 1.7 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 20 kHz |
|
0.017 % |
+ 1.7 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
20 kHz to 50 kHz |
|
0.038 % |
+ 1.7 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
50 kHz to 100 kHz |
|
0.13 % |
+ 1.9 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
100 kHz to 300 kHz |
|
0.37 % |
+ 12 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
1.2 % |
+ 11 mV |
|
|
|
|
|||||||||||||
|
|
120 |
mV |
to |
1.2 |
V |
1 Hz to 40 Hz |
|
0.028 % |
+ 32 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
40 Hz to 1,000 Hz |
|
0.01 % |
+ 22 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 20 kHz |
|
0.01 % |
+ 22 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
20 kHz to 50 kHz |
|
0.035 % |
+ 22 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
50 kHz to 100 kHz |
|
0.093 % |
+ 23 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
100 kHz to 300 kHz |
|
0.35 % |
+ 120 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
1.2 % |
+ 80 mV |
|
|
|
|
|||||||||||||
|
|
1.2 |
V |
to |
12 |
V |
1 Hz to 40 Hz |
|
0.028 % |
+ 320 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
40 Hz to 1,000 Hz |
|
0.0093 % |
+ 220 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 20 kHz |
|
0.0093 % |
+ 220 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
20 kHz to 50 kHz |
|
0.035 % |
+ 220 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
50 kHz to 100 kHz |
|
0.093 % |
+ 230 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
100 kHz to 300 kHz |
|
0.35 % |
+ 1.2 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
1.2 % |
+ 0.77 mV |
|
|
|
|
|||||||||||||
|
|
12 |
V |
to |
120 |
V |
1 Hz to 40 Hz |
|
0.036 % |
+ 3.8 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
40 Hz to 1,000 Hz |
|
0.024 % |
+ 2.2 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 20 kHz |
|
0.024 % |
+ 2.2 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
20 kHz to 50 kHz |
|
0.041 % |
+ 2.2 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
50 kHz to 100 kHz |
|
0.14 % |
+ 2.3 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
100 kHz to 300 kHz |
|
0.47 % |
+ 12 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
1.2 % |
+ 16 mV |
|
|
|
|
|||||||||||||
|
|
120 |
V |
to |
700 |
V |
1 Hz to 40 Hz |
|
0.047 % |
+ 46 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
40 Hz to 1,000 Hz |
|
0.047 % |
+ 23 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 20 kHz |
|
0.047 % |
+ 23 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
20 kHz to 50 kHz |
|
0.014 % |
+ 23 mV |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
50 kHz to 100 kHz |
|
0.24 % |
+ 23 mV |
|
|
|
|
|||||||||||||
|
|
1,100 |
V |
to |
5,000 |
V |
40 Hz to 100 Hz |
|
2 % |
|
Measurement by a standard multimeter with a HV probe |
KP 8/E |
1 |
|
|||||||||||||
7 |
Alternating-current voltage, peak and peak-to-peak value of periodic non-harmonic signals / sources of non-harmonic and pulse signals, pulse testers |
|
|
|
|
|
Bandwidth: |
Impedance: |
|
|
Measurement by a standard oscilloscope5 |
KP 8/E |
1 |
|
|||||||||||||
|
|
10 |
mV |
to |
5 |
V |
1 Hz to 300 kHz, |
Z = 50 W |
1.8 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 kHz to 2 GHz |
|
4.0 % |
|
|
|
|
|
|||||||||||||
|
|
5 |
V |
to |
38 |
V |
1 Hz to 300 kHz, |
Z = 50 W |
2.9 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 GHz |
|
4.5 % |
|
|
|
|
|
|||||||||||||
|
|
10 |
mV |
to |
40 |
V |
1 Hz to 300 kHz, |
Z = 1 MW |
1.8 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
300 kHz to 400 MHz |
|
4.0 % |
|
|
|
|
|
|||||||||||||
|
|
40 |
V |
to |
400 |
V |
1 Hz to 400 MHz |
Z = 10 MW |
4.5 % |
|
|
|
|
|
|||||||||||||
|
|
40 |
V |
to |
1,000 |
V |
1 Hz to 400 MHz |
Z = 66.7 MW |
4.5 % |
|
|
|
|
|
|||||||||||||
8 |
Pulse amplitude, peak and peak-to-peak value of voltage / sources of non-harmonic and pulse signals, pulse testers |
10 |
mV |
to |
40 |
V |
Z = 1 MW |
rise time >1.3 ns |
4.0 % |
|
Measurement by a standard oscilloscope with a voltage probe |
KP 8/E |
1 |
|
|||||||||||||
|
|
10 |
V |
to |
5,000 |
V |
Z = 50 kW |
rise time >140 ns |
5.6 % |
|
|
|
|
|
|||||||||||||
|
|
10 |
V |
to |
5,000 |
V |
Z = 1 kW |
rise time >4.2 ns |
5.0 % |
|
|
|
|
|
|||||||||||||
|
|
1 |
V |
to |
5,000 |
V |
Z = 50 W |
rise time >3.5 ns |
4.9 % |
|
|
|
|
|
|||||||||||||
9 |
Alternating current, peak and peak-to-peak value / sources of non-harmonic and pulse signals, pulse testers |
100 |
mA |
to |
1.5 |
A |
1 Hz to 20 kHz |
|
2.7 % |
|
Measurement by a standard oscilloscope with a shunt or clamp probe |
KP 8/E |
1 |
|
|||||||||||||
|
|
1.5 |
A |
to |
15 |
A |
1 Hz to 20 kHz |
|
2.4 % |
|
|
|
|
|
|||||||||||||
|
|
1 |
A |
to |
2,500 |
A |
|
rise time >140 ns |
4.9 % |
|
Measurement by a standard oscilloscope with a current shunt |
|
|
|
|||||||||||||
10 |
Alternating current / digital, analogue multimeters and ammeters, testing and inspection instruments and their parts, current probes |
1 |
mA |
to |
10 |
mA |
10 Hz to 1 kHz |
|
|
0.015 mA |
Current generation by a standard calibrator |
KP 1/E, KP 2/E, KP 3/E |
1 |
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
|
0.02 mA |
|
|
|
|
|||||||||||||
|
|
10 |
mA |
to |
50 |
mA |
10 Hz to 1 kHz |
|
|
0.025 mA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
|
0.035 mA |
|
|
|
|
|||||||||||||
|
|
50 |
mA |
to |
100 |
mA |
10 Hz to 1 kHz |
|
|
0.035 mA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
|
0.06 mA |
|
|
|
|
|||||||||||||
|
|
100 |
mA |
to |
200 |
mA |
10 Hz to 1 kHz |
|
0.035 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
0.06 % |
|
|
|
|
|
|||||||||||||
|
|
200 |
mA |
to |
400 |
mA |
10 Hz to 1 kHz |
|
0.071 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
0.083 % |
|
|
|
|
|
|||||||||||||
|
|
400 |
mA |
to |
800 |
mA |
10 Hz to 1 kHz |
|
0.045 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
0.06 % |
|
|
|
|
|
|||||||||||||
|
|
800 |
mA |
to |
2 |
mA |
10 Hz to 1 kHz |
|
0.03 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
0.045 % |
|
|
|
|
|
|||||||||||||
|
|
2 |
mA |
to |
3 |
mA |
10 Hz to 1 kHz |
|
0.075 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
0.085 % |
|
|
|
|
|
|||||||||||||
|
|
3 |
mA |
to |
4 |
mA |
10 Hz to 1 kHz |
|
0.055 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
0.065 % |
|
|
|
|
|
|||||||||||||
|
|
4 |
mA |
to |
8 |
mA |
10 Hz to 1 kHz |
|
0.045 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
0.06 % |
|
|
|
|
|
|||||||||||||
|
|
8 |
mA |
to |
20 |
mA |
10 Hz to 1 kHz |
|
0.033 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
0.045 % |
|
|
|
|
|
|||||||||||||
|
|
20 |
mA |
to |
30 |
mA |
10 Hz to 1 kHz |
|
0.075 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
0.085 % |
|
|
|
|
|
|||||||||||||
|
|
30 |
mA |
to |
40 |
mA |
10 Hz to 1 kHz |
|
0.055 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
0.065 % |
|
|
|
|
|
|||||||||||||
|
|
40 |
mA |
to |
80 |
mA |
10 Hz to 1 kHz |
|
0.045 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
0.055 % |
|
|
|
|
|
|||||||||||||
|
|
80 |
mA |
to |
200 |
mA |
10 Hz to 1 kHz |
|
0.03 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
0.045 % |
|
|
|
|
|
|||||||||||||
|
|
200 |
mA |
to |
300 |
mA |
10 Hz to 1 kHz |
|
0.15 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
2 % |
|
|
|
|
|
|||||||||||||
|
|
300 |
mA |
to |
800 |
mA |
10 Hz to 1 kHz |
|
0.1 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
0.15 % |
|
|
|
|
|
|||||||||||||
|
|
800 |
mA |
to |
2,000 |
mA |
10 Hz to 1 kHz |
|
0.06 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
0.11 % |
|
|
|
|
|
|||||||||||||
|
|
2 |
A |
to |
4 |
A |
10 Hz to 1 kHz |
|
0.1 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
0.2 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
5 kHz to 10 kHz |
|
0.5 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
10 kHz to 20 kHz |
|
2 % |
|
|
|
|
|
|||||||||||||
|
|
4 |
A |
to |
6 |
A |
10 Hz to 1 kHz |
|
0.07 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
0.14 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
5 kHz to 10 kHz |
|
0.38 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
10 kHz to 20 kHz |
|
1.5 % |
|
|
|
|
|
|||||||||||||
|
|
6 |
A |
to |
11 |
A |
10 Hz to 1 kHz |
|
0.07 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
0.13 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
5 kHz to 10 kHz |
|
0.35 % |
|
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
10 kHz to 20 kHz |
|
1.3 % |
|
|
|
|
|
|||||||||||||
|
|
11 |
A |
to |
20 |
A |
40 Hz to 400 Hz |
|
0.35 % |
|
|
|
|
|
|||||||||||||
|
|
20 |
A |
to |
30 |
A |
40 Hz to 300 Hz |
|
0.1 % |
|
Comparison with current measured by a standard shunt with multimeter |
KP 1/E, KP 2/E, KP 3/E |
1 |
|
|||||||||||||
|
|
30 |
A |
to |
50 |
A |
40 Hz to 250 Hz |
|
1.2 % |
|
|
|
|
|
|||||||||||||
|
|
50 |
A |
to |
100 |
A |
40 Hz to 160 Hz |
|
2 % |
|
|
|
|
|
|||||||||||||
|
Clamp ammeters and clamp current probes |
1 |
A |
to |
5 |
A |
40 Hz to 300 Hz |
|
0.25 % |
|
Direct generation of current with a calibrator |
KP 1/E |
|
|
|||||||||||||
|
|
5 |
A |
to |
20 |
A |
40 Hz to 300 Hz |
|
0.16 % |
|
Simulation of reference current using calibrator and current coil |
|
|
|
|||||||||||||
|
|
20 |
A |
to |
30 |
A |
40 Hz to 300 Hz |
|
0.33 % |
|
|
|
|
|
|||||||||||||
|
|
30 |
A |
to |
100 |
A |
40 Hz to 300 Hz |
|
0.25 % |
|
|
|
|
|
|||||||||||||
|
|
100 |
A |
to |
250 |
A |
40 Hz to 300 Hz |
|
0.35 % |
|
|
|
|
|
|||||||||||||
|
|
250 |
A |
to |
750 |
A |
40 Hz to 300 Hz |
|
0.2 % |
|
|
|
|
|
|||||||||||||
11 |
Alternating current / current sources |
1 |
mA |
to |
120 |
mA |
10 Hz to 20 Hz |
|
0.46 % |
+ 35 nA |
Direct measurement by a standard multimeter |
KP 8/E |
1 |
|
|||||||||||||
|
|
|
|
|
|
|
20 Hz to 45 Hz |
|
0.17 % |
+ 35 nA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
45 Hz to 100 Hz |
|
0.07 % |
+ 35 nA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
100 Hz to 1,000 Hz |
|
0.071 % |
+ 35 nA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
0.061 % |
+ 35 nA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
5 kHz to 20 kHz |
|
0.061 % |
+ 35 nA |
|
|
|
|
|||||||||||||
|
|
120 |
mA |
to |
1.2 |
mA |
10 Hz to 20 Hz |
|
0.46 % |
+ 230 nA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
20 Hz to 45 Hz |
|
0.17 % |
+ 230 nA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
45 Hz to 100 Hz |
|
0.07 % |
+ 230 nA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
100 Hz to 1,000 Hz |
|
0.037 % |
+ 230 nA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
0.044 % |
+ 230 nA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
5 kHz to 20 kHz |
|
0.044 % |
+ 230 nA |
|
|
|
|
|||||||||||||
|
|
1.2 |
mA |
to |
12 |
mA |
10 Hz to 20 Hz |
|
0.46 % |
+ 2.3 mA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
20 Hz to 45 Hz |
|
0.17 % |
+ 2.3 mA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
45 Hz to 100 Hz |
|
0.07 % |
+ 2.3 mA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
100 Hz to 1,000 Hz |
|
0.037 % |
+ 2.3 mA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
0.041 % |
+ 2.3 mA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
5 kHz to 20 kHz |
|
0.041 % |
+ 2.3 mA |
|
|
|
|
|||||||||||||
|
|
12 |
mA |
to |
120 |
mA |
10 Hz to 20 Hz |
|
0.46 % |
+ 23 mA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
20 Hz to 45 Hz |
|
0.17 % |
+ 23 mA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
45 Hz to 100 Hz |
|
0.07 % |
+ 23 mA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
100 Hz to 1,000 Hz |
|
0.037 % |
+ 23 mA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
0.04 % |
+ 23 mA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
5 kHz to 20 kHz |
|
0.04 % |
+ 23 mA |
|
|
|
|
|||||||||||||
|
|
120 |
mA |
to |
1 |
A |
10 Hz to 20 Hz |
|
0.46 % |
+ 230 mA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
20 Hz to 45 Hz |
|
0.17 % |
+ 230 mA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
45 Hz to 100 Hz |
|
0.07 % |
+ 230 mA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
100 Hz to 1,000 Hz |
|
0.12 % |
+ 230 mA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 5 kHz |
|
0.13 % |
+ 230 mA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
5 kHz to 20 kHz |
|
0.13 % |
+ 230 mA |
|
|
|
|
|||||||||||||
|
|
1.2 |
A |
to |
12 |
A |
40 Hz to 1 kHz |
|
0.14 % |
+ 330 mA |
Indirect measurement by a multimeter with a standard shunt |
KP 8/E |
1 |
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 20 kHz |
|
0.15 % |
+ 41 mA |
|
|
|
|
|||||||||||||
|
|
12 |
A |
to |
20 |
A |
40 Hz to 1 kHz |
|
0.14 % |
+ 420 mA |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1 kHz to 20 kHz |
|
0.15 % |
+ 490 mA |
|
|
|
|
|||||||||||||
|
|
20 |
A |
to |
50 |
A |
40 Hz to 250 Hz |
|
1 % |
+ 44 mA |
|
|
|
|
|||||||||||||
|
|
50 |
A |
to |
100 |
A |
40 Hz to 250 Hz |
|
2 % |
+ 48 mA |
|
|
|
|
|||||||||||||
|
|
1 |
A |
to |
40 |
A |
40 Hz to 300 Hz |
|
1.2 % |
+ 0.5 A |
Measurement by a standard clamp ammeter |
KP 8/E |
1 |
|
|||||||||||||
|
|
40 |
A |
to |
400 |
A |
40 Hz to 300 Hz |
|
0.6 % |
+ 5 A |
|
|
|
|
|||||||||||||
|
|
400 |
A |
to |
1,500 |
A |
40 Hz to 150 Hz |
|
0.6 % |
+ 23 A |
|
|
|
|
|||||||||||||
12* |
DC resistance / ohmmeters, testing and inspection instruments and their parts. |
|
|
|
0 |
W |
|
|
|
20 mW |
Direct measurement of fixed resistance standards |
KP 1/E, KP 2/E |
1 |
|
|||||||||||||
|
|
|
|
|
0.1 |
mW |
|
|
|
6 nW |
|
|
|
|
|||||||||||||
|
|
|
|
|
1 |
mW |
|
|
|
50 nW |
|
|
|
|
|||||||||||||
|
|
|
|
|
0.01 |
W |
|
|
|
0.3 mW |
|
|
|
|
|||||||||||||
|
|
|
|
|
0.1 |
W |
|
|
|
2 mW |
|
|
|
|
|||||||||||||
|
|
|
|
|
1 |
W |
|
|
|
12 mW |
|
|
|
|
|||||||||||||
|
|
|
|
|
10 |
W |
|
|
|
200 mW |
|
|
|
|
|||||||||||||
|
|
|
|
|
100 |
W |
|
|
|
1.2 mW |
|
|
|
|
|||||||||||||
|
|
|
|
|
1 |
kW |
|
|
|
14 mW |
|
|
|
|
|||||||||||||
|
|
|
|
|
10 |
kW |
|
|
|
0.14 W |
|
|
|
|
|||||||||||||
|
|
|
|
|
100 |
kW |
|
|
|
1.7 W |
|
|
|
|
|||||||||||||
|
|
|
|
|
1 |
MW |
|
|
|
30 W |
|
|
|
|
|||||||||||||
|
|
|
|
|
10 |
MW |
|
|
|
60 W |
|
|
|
|
|||||||||||||
|
|
|
|
|
100 |
MW |
|
|
|
9 kW |
|
|
|
|
|||||||||||||
|
|
|
|
|
1 |
GW |
|
|
|
0.4 MW |
|
|
|
|
|||||||||||||
|
|
|
|
|
10 |
GW |
|
|
|
0.4 GW |
|
|
|
|
|||||||||||||
|
|
|
|
|
100 |
GW |
|
|
|
5 GW |
|
|
|
|
|||||||||||||
|
|
|
|
|
1 |
TW |
|
|
|
100 GW |
|
|
|
|
|||||||||||||
13* |
DC resistance/ standard resistors, shunts |
|
|
|
0.1 |
mW |
|
|
|
10 nW |
Comparison by ratio method – measurement of voltage drop at constant measuring current |
KP 6/E |
1 |
|
|||||||||||||
|
|
|
|
|
1 |
mW |
|
|
|
70 nW |
|
|
|
|
|||||||||||||
|
|
|
|
|
0.01 |
W |
|
|
|
0.4 mW |
|
|
|
|
|||||||||||||
|
|
|
|
|
0.1 |
W |
|
|
|
3 mW |
|
|
|
|
|||||||||||||
|
|
|
|
|
1 |
W |
|
|
|
20 mW |
|
|
|
|
|||||||||||||
|
|
|
|
|
10 |
W |
|
|
|
250 mW |
|
|
|
|
|||||||||||||
|
|
|
|
|
100 |
W |
|
|
|
1.5 mW |
|
|
|
|
|||||||||||||
|
|
|
|
|
1 |
kW |
|
|
|
40 mW |
|
|
|
|
|||||||||||||
|
|
|
|
|
10 |
kW |
|
|
|
0.4 W |
|
|
|
|
|||||||||||||
|
|
|
|
|
100 |
kW |
|
|
|
7 W |
|
|
|
|
|||||||||||||
|
|
|
|
|
1 |
MW |
|
|
|
40 W |
Comparison by ratio method – measurement of standard and calibrated resistance |
|
|
|
|||||||||||||
|
|
|
|
|
10 |
MW |
|
|
|
700 W |
|
|
|
|
|||||||||||||
|
|
|
|
|
100 |
MW |
|
|
|
12 kW |
|
|
|
|
|||||||||||||
|
|
|
|
|
1 |
GW |
|
|
|
0.6 MW |
|
|
|
|
|||||||||||||
14* |
DC resistance / resistors, calibrators, simulators, testing and inspection instruments and their parts. |
0 |
W |
to |
12 |
W |
|
|
0.002 % |
+ 78 mW |
Direct measurement by a standard multimeter |
KP 8/E |
1 |
|
|||||||||||||
|
|
12 |
W |
to |
120 |
W |
|
|
0.0016 % |
+ 570 mW |
|
|
|
|
|||||||||||||
|
|
0.12 |
kW |
to |
1.2 |
kW |
|
|
0.0014 % |
+ 540 mW |
|
|
|
|
|||||||||||||
|
|
1.2 |
kW |
to |
12 |
kW |
|
|
0.0014 % |
+ 5.4 mW |
|
|
|
|
|||||||||||||
|
|
12 |
kW |
to |
120 |
kW |
|
|
0.0016 % |
+ 45 mW |
|
|
|
|
|||||||||||||
|
|
0.12 |
MW |
to |
1.2 |
MW |
|
|
0.0026 % |
+ 1.9 W |
|
|
|
|
|||||||||||||
|
|
1.2 |
MW |
to |
12 |
MW |
|
|
0.0068 % |
+ 110 W |
|
|
|
|
|||||||||||||
|
|
12 |
MW |
to |
120 |
MW |
|
|
0.058 % |
+ 1.3 kW |
|
|
|
|
|||||||||||||
|
|
0.12 |
GW |
to |
1.2 |
GW |
|
|
0.57 % |
+ 0.35 MW |
|
|
|
|
|||||||||||||
15 |
HF level (voltage) / HF voltmeters, measuring receivers, synthesizers, oscilloscopes, spectral and circuit analyzers, radiocommunication and radio navigation testers, simulators and imitators (ATC, DME, TACAN, TCAS, VOR-ILS) |
20
|
dBmV µV |
to |
47
|
dBmV mV) |
0.1 MHz to 1.4 GHz |
|
4 % |
|
Indirect comparison with standard HF wattmeter |
KP 13/E, KP 18/E |
1 |
||||||||||||||
|
|
|
|
|
|
|
1.4 GHz to 2.08 GHz |
|
3.5 % |
|
|
|
|
||||||||||||||
|
|
47
|
dBmV
|
to |
67
|
dBmV
|
10 MHz to 30 MHz |
|
4 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
30 MHz to 2.08 GHz |
|
3.5 % |
|
|
|
|
||||||||||||||
|
|
67
|
dBmV
|
to |
87
|
dBmV
|
10 MHz to 30 MHz |
|
4 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
30 MHz to 2.08 GHz |
|
3.5 % |
|
|
|
|
||||||||||||||
|
|
87
|
dBmV
|
to |
117
|
dBmV
|
100 kHz to 300 kHz |
|
17 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
300 kHz to 600 kHz |
|
14 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
600 kHz to 1 MHz |
|
8.5 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 MHz to 1.3 GHz |
|
2.6 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1.3 GHz to 2.08 GHz |
|
3.2 % |
|
|
|
|
||||||||||||||
|
|
117
|
dBmV
|
to |
137
|
dBmV
|
100 kHz to 300 kHz |
|
17 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
300 kHz to 600 kHz |
|
14 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
600 kHz to 1 MHz |
|
8.5 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 MHz to 1.3 GHz |
|
3.5 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1.3 GHz to 2.08 GHz |
|
4 % |
|
|
|
|
||||||||||||||
|
|
137
|
dBmV
|
to |
147
|
dBmV
|
100 kHz to 2.08 GHz |
|
2.9 % |
|
|
|
|
||||||||||||||
|
|
147
|
dBmV
|
to |
152
|
dBmV
|
100 kHz to 2.08 GHz |
|
3.7 % |
|
|
|
|
||||||||||||||
16 |
HF level (voltage) / HF generators, circuit analyzers, radiocommunication and radio navigation testers, simulators and imitators (ATC, DME, TACAN, TCAS, VOR-ILS) |
20
|
dBmV µV |
to |
47
|
dBmV mV) |
0.1 MHz to 1.4 GHz |
|
4 % |
|
Indirect measurement of HF voltage - by measuring HF power at impedance 50 W6
|
KP 11/E, KP 12/E, |
1 |
||||||||||||||
|
|
|
|
|
|
|
1.4 GHz to 2.08 GHz |
|
3.5 % |
|
|
|
|
||||||||||||||
|
|
47
|
dBmV
|
to |
67
|
dBmV
|
10 MHz to 30 MHz |
|
4 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
30 MHz to 2.08 GHz |
|
3.5 % |
|
|
|
|
||||||||||||||
|
|
67
|
dBmV
|
to |
87
|
dBmV
|
10 MHz to 30 MHz |
|
4 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
30 MHz to 2.08 GHz |
|
3.5 % |
|
|
|
|
||||||||||||||
|
|
87
|
dBmV
|
to |
117
|
dBmV
|
100 kHz to 300 kHz |
|
17 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
300 kHz to 600 kHz |
|
14 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
600 kHz to 1 MHz |
|
8.5 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 MHz to 1.3 GHz |
|
2.6 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1.3 GHz to 2.08 GHz |
|
3.2 % |
|
|
|
|
||||||||||||||
|
|
117
|
dBmV
|
to |
137
|
dBmV
|
100 kHz to 300 kHz |
|
17 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
300 kHz to 600 kHz |
|
14 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
600 kHz to 1 MHz |
|
8.5 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 MHz to 1.3 GHz |
|
3.5 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1.3 GHz to 2.08 GHz |
|
4 % |
|
|
|
|
||||||||||||||
|
|
137
|
dBmV
|
to |
147
|
dBmV
|
100 kHz to 2.08 GHz |
|
2.9 % |
|
|
|
|
||||||||||||||
|
|
147
|
dBmV
|
to |
152
|
dBmV
|
100 kHz to 2.08 GHz |
|
3.7 % |
|
|
|
|
||||||||||||||
17* |
HF level (power) / oscillators, circuit analyzers, testers, radiocommunication and radio navigation testers, simulators and imitators (ATC, DME, TACAN, TCAS, VOR-ILS) |
|
|
|
|
|
|
|
|
|
Direct measurement by a standard measuring receiver |
KP 15/E |
1 |
||||||||||||||
|
|
1
|
pW
|
to |
2.5
|
pW
|
10 MHz to 20 MHz |
|
8.7 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
20 MHz to 50 MHz |
|
6.5 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
50 MHz to 7 GHz |
|
6.2 % |
|
|
|
|
||||||||||||||
|
|
2.5
|
pW
|
to |
800
|
pW
|
10 MHz to 20 MHz |
|
7.8 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
20 MHz to 50 MHz |
|
5.2 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
50 MHz to 7 GHz |
|
4.8 % |
|
|
|
|
||||||||||||||
|
|
800
|
pW
|
to |
2
|
nW
|
10 MHz to 20 MHz |
|
8.5 % |
|
Measurement by a standard HF wattmeter |
KP 15/E |
|
||||||||||||||
|
|
|
|
|
|
|
20 MHz to 50 MHz |
|
6.2 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
50 MHz to 7 GHz |
|
5.5 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
7 GHz to 15 GHz |
|
6.0 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
15 GHz to 18 GHz |
|
6.6 % |
|
|
|
|
||||||||||||||
|
|
2
|
nW
|
to |
10
|
μW
|
10 MHz to 20 MHz |
|
7.5 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
20 MHz to 50 MHz |
|
4.8 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
50 MHz to 7 GHz |
|
3.9 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
7 GHz to 15 GHz |
|
4.5 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
15 GHz to 18 GHz |
|
5.4 % |
|
|
|
|
||||||||||||||
|
|
10
|
μW
|
to |
100
|
μW
|
100 kHz to 300 kHz |
|
7.3 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
4.0 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 MHz to 50 MHz |
|
3.6 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
50 MHz to 1 GHz |
|
3.0 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 GHz to 2 GHz |
|
3.7 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
2 GHz to 15 GHz |
|
4.8 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
15 GHz to 18 GHz |
|
5.1 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
18 GHz to 20 GHz |
|
5.5 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
20 GHz to 26.5 GHz |
|
4.6 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
26.5 GHz to 40.8 GHz |
|
6.0 % |
|
|
|
|
||||||||||||||
|
|
100
|
μW
|
to |
10
|
mW
|
100 kHz to 300 kHz |
|
7.7 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
4.1 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 MHz to 50 MHz |
|
3.4 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
50 MHz to 1 GHz |
|
2.6 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 GHz to 2 GHz |
|
3.7 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
2 GHz to 15 GHz |
|
4.6 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
15 GHz to 18 GHz |
|
4.9 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
18 GHz to 20 GHz |
|
5.3 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
20 GHz to 26.5 GHz |
|
4.3 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
26.5 GHz to 40.8 GHz |
|
5.6 % |
|
|
|
|
||||||||||||||
|
|
10
|
mW
|
to |
100
|
mW
|
|
|
|
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
100 kHz to 300 kHz |
|
6.2 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
6.2 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 MHz to 10 MHz |
|
5.4 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
10 MHz to 50 MHz |
|
4.6 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
50 MHz to 1 GHz |
|
4.4 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 GHz to 15 GHz |
|
5.8 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
15 GHz to 18 GHz |
|
6.0 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
18 GHz to 20 GHz |
|
6.4 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
20 GHz to 26.5 GHz |
|
5.5 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
26.5 GHz to 40.8 GHz |
|
6.5 % |
|
|
|
|
||||||||||||||
|
|
100 |
mW |
to |
3 |
W |
|
|
|
|
|
|
|
||||||||||||||
|
|
(+20 |
dBm |
to |
+33 |
dBm) |
100 kHz to 300 kHz |
|
6.0 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
6.0 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 MHz to 10 MHz |
|
5.2 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
10 MHz to 500 MHz |
|
4.8 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
500 MHz to 1 GHz |
|
5.9 % |
|
|
|
|
||||||||||||||
|
|
3
|
mW
|
to |
10
|
W
|
1 GHz to 4.2 GHz |
|
6.7 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
100 kHz to 300 kHz |
|
7.3 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
7.3 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 MHz to 10 MHz |
|
6.6 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
10 MHz to 500 MHz |
|
6.3 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
500 MHz to 1 GHz |
|
7.1 % |
|
|
|
|
||||||||||||||
|
|
10
|
W
|
to |
25
|
W
|
1 GHz to 4.2 GHz |
|
7.6 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
100 kHz to 300 kHz |
|
7.4 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
7.4 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 MHz to 10 MHz |
|
6.7 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
10 MHz to 500 MHz |
|
6.2 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
500 MHz to 1 GHz |
|
6.9 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 GHz to 2 GHz |
|
8.0 % |
|
|
|
|
||||||||||||||
|
|
10
|
W
|
to |
25
|
W
|
2 GHz to 4.2 GHz |
|
9.5 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 GHz to 2 GHz |
|
8.0 % |
|
|
|
|
||||||||||||||
18* |
HF level (power) / HF wattmeters, selective HF meters, spectral and circuit analyzers, testers, radiocommunication and radio navigation testers, simulators and imitators (ATC, DME, TACAN, TCAS, VOR-ILS) |
|
|
|
|
|
|
|
|
|
Comparison with a standard HF wattmeter |
KP 14/E |
1 |
||||||||||||||
|
|
800
|
pW
|
to |
2
|
nW
|
10 MHz to 20 MHz |
|
8.5 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
20 MHz to 50 MHz |
|
6.2 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
50 MHz to 7 GHz |
|
5.5 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
7 GHz to 15 GHz |
|
6.0 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
15 GHz to 18 GHz |
|
6.6 % |
|
|
|
|
||||||||||||||
|
|
2
|
nW
|
to |
10
|
μW
|
10 MHz to 20 MHz |
|
7.5 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
20 MHz to 50 MHz |
|
4.8 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
50 MHz to 7 GHz |
|
3.9 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
7 GHz to 15 GHz |
|
4.5 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
15 GHz to 18 GHz |
|
5.4 % |
|
|
|
|
||||||||||||||
|
|
10
|
μW
|
to |
100
|
μW
|
100 kHz to 300 kHz |
|
7.3 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
4.0 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 MHz to 50 MHz |
|
3.6 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
50 MHz to 1 GHz |
|
3.0 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 GHz to 2 GHz |
|
3.7 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
2 GHz to 15 GHz |
|
4.8 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
15 GHz to 18 GHz |
|
5.1 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
18 GHz to 20 GHz |
|
5.5 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
20 GHz to 26.5 GHz |
|
4.6 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
26.5 GHz to 40.8 GHz |
|
6.0 % |
|
|
|
|
||||||||||||||
|
|
100
|
μW
|
to |
10
|
mW
|
100 kHz to 300 kHz |
|
7.7 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
4.1 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 MHz to 50 MHz |
|
3.4 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
50 MHz to 1 GHz |
|
2.6 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 GHz to 2 GHz |
|
3.7 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
2 GHz to 15 GHz |
|
4.6 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
15 GHz to 18 GHz |
|
4.9 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
18 GHz to 20 GHz |
|
5.3 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
20 GHz to 26.5 GHz |
|
4.3 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
26.5 GHz to 40.8 GHz |
|
5.6 % |
|
|
|
|
||||||||||||||
|
|
10
|
mW
|
to |
100
|
mW
|
|
|
|
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
100 kHz to 300 kHz |
|
6.2 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
6.2 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 MHz to 10 MHz |
|
5.4 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
10 MHz to 50 MHz |
|
4.6 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
50 MHz to 1 GHz |
|
4.4 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 GHz to 15 GHz |
|
5.8 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
15 GHz to 18 GHz |
|
6.0 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
18 GHz to 20 GHz |
|
6.4 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
20 GHz to 26.5 GHz |
|
5.5 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
26.5 GHz to 40.8 GHz |
|
6.5 % |
|
|
|
|
||||||||||||||
|
|
100 |
mW |
to |
2 |
W |
|
|
|
|
|
|
|
||||||||||||||
|
|
(+20 |
dBm |
to |
+33 |
dBm) |
100 kHz to 300 kHz |
|
6.0 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
6.0 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 MHz to 10 MHz |
|
5.2 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
10 MHz to 500 MHz |
|
4.8 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
500 MHz to 1 GHz |
|
5.9 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 GHz to 4.2 GHz |
|
6.7 % |
|
|
|
|
||||||||||||||
|
|
2
|
mW
|
to |
10
|
W
|
100 kHz to 300 kHz |
|
7.3 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
7.3 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 MHz to 10 MHz |
|
6.6 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
10 MHz to 500 MHz |
|
6.3 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
500 MHz to 1 GHz |
|
7.1 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 GHz to 4.2 GHz |
|
7.6 % |
|
|
|
|
||||||||||||||
|
|
10
|
W
|
to |
25
|
W
|
100 kHz to 300 kHz |
|
7.4 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
300 kHz to 1 MHz |
|
7.4 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 MHz to 10 MHz |
|
6.7 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
10 MHz to 500 MHz |
|
6.2 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
500 MHz to 1 GHz |
|
6.9 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
1 GHz to 2 GHz |
|
8.0 % |
|
|
|
|
||||||||||||||
|
|
|
|
|
|
|
2 GHz to 4.2 GHz |
|
9.5 % |
|
|
|
|
||||||||||||||
19* |
HF attenuation / attenuators, attenuation pads, selective HF meters, spectral and circuit analyzers, testers, radiocommunication and radio navigation testers, simulators, imitators (ATC, DME, TACAN, TCAS, VOR-ILS) |
0 |
dBm |
to |
10 |
dBm |
100 kHz to 400 kHz |
|
|
0.048 dB |
Indirect measurement power method |
KP 16/E |
1 |
|
|||||||||||||
|
|
|
|
|
|
|
400 kHz to 1.3 GHz |
|
|
0.048 dB |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1.3 GHz to 3 GHz |
|
|
0.048 dB |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
3 GHz to 6 GHz |
|
|
0.059 dB |
|
|
|
|
|||||||||||||
|
|
10 |
dBm |
to |
20 |
dBm |
100 kHz to 400 kHz |
|
|
0.064 dB |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
400 kHz to 1.3 GHz |
|
|
0.067 dB |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1.3 GHz to 3 GHz |
|
|
0.062 dB |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
3 GHz to 6 GHz |
|
|
0.076 dB |
|
|
|
|
|||||||||||||
|
|
20 |
dBm |
to |
30 |
dBm |
100 kHz to 400 kHz |
|
|
0.11 dB |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
400 kHz to 1.3 GHz |
|
|
0.11 dB |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
1.3 GHz to 3 GHz |
|
|
0.11 dB |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
3 GHz to 6 GHz |
|
|
0.11 dB |
|
|
|
|
|||||||||||||
|
|
30 |
dBm |
to |
40 |
dBm |
100 kHz to 400 kHz |
|
|
0.15 dB |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
400 kHz to 3 GHz |
|
|
0.15 dB |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
3 GHz to 6 GHz |
|
|
0.17 dB |
|
|
|
|
|||||||||||||
|
|
40 |
dBm |
to |
60 |
dBm |
100 kHz to 3 GHz |
|
|
0.18 dB |
|
|
|
|
|||||||||||||
|
|
|
|
|
|
|
3 GHz to 6 GHz |
|
|
0.19 dB |
|
|
|
|
|||||||||||||
|
|
60 |
dBm |
to |
65 |
dBm |
100 kHz to 6 GHz |
|
|
0.25 dB |
|
|
|
|
19* |
HF attenuation / attenuators, attenuation pads, selective HF meters, spectral and circuit analyzers, testers, radiocommunication and radio navigation testers, simulators, imitators (ATC, DME, TACAN, TCAS, VOR-ILS) |
0 |
dBm |
to |
10 |
dBm |
100 kHz to 400 kHz |
|
|
0.048 dB |
Indirect measurement power method |
KP 16/E |
1 |
||||
|
|
|
|
|
|
|
400 kHz to 1.3 GHz |
|
|
0.048 dB |
|
|
|
||||
|
|
|
|
|
|
|
1.3 GHz to 3 GHz |
|
|
0.048 dB |
|
|
|
||||
|
|
|
|
|
|
|
3 GHz to 6 GHz |
|
|
0.059 dB |
|
|
|
||||
|
|
10 |
dBm |
to |
20 |
dBm |
100 kHz to 400 kHz |
|
|
0.064 dB |
|
|
|
||||
|
|
|
|
|
|
|
400 kHz to 1.3 GHz |
|
|
0.067 dB |
|
|
|
||||
|
|
|
|
|
|
|
1.3 GHz to 3 GHz |
|
|
0.062 dB |
|
|
|
||||
|
|
|
|
|
|
|
3 GHz to 6 GHz |
|
|
0.076 dB |
|
|
|
||||
|
|
20 |
dBm |
to |
30 |
dBm |
100 kHz to 400 kHz |
|
|
0.11 dB |
|
|
|
||||
|
|
|
|
|
|
|
400 kHz to 1.3 GHz |
|
|
0.11 dB |
|
|
|
||||
|
|
|
|
|
|
|
1.3 GHz to 3 GHz |
|
|
0.11 dB |
|
|
|
||||
|
|
|
|
|
|
|
3 GHz to 6 GHz |
|
|
0.11 dB |
|
|
|
||||
|
|
30 |
dBm |
to |
40 |
dBm |
100 kHz to 400 kHz |
|
|
0.15 dB |
|
|
|
||||
|
|
|
|
|
|
|
400 kHz to 3 GHz |
|
|
0.15 dB |
|
|
|
||||
|
|
|
|
|
|
|
3 GHz to 6 GHz |
|
|
0.17 dB |
|
|
|
||||
|
|
40 |
dBm |
to |
60 |
dBm |
100 kHz to 3 GHz |
|
|
0.18 dB |
|
|
|
||||
|
|
|
|
|
|
|
3 GHz to 6 GHz |
|
|
0.19 dB |
|
|
|
||||
|
|
60 |
dBm |
to |
65 |
dBm |
100 kHz to 6 GHz |
|
|
0.25 dB |
|
|
|
||||
20* |
Modulation swing of frequency modulation / generators, modulation meters, selective HF meters, spectral analyzers, radiocommunication and radio navigation testers, simulators and imitators (ATC, DME, TACAN, TCAS, VOR-ILS) |
|
|
|
|
|
Modulation frequency |
Carrier frequency |
|
|
Direct measurement or comparison with a modulation analyzer |
KP 21/E |
1 |
||||
|
|
0.1 |
kHz |
to |
0.2 |
kHz |
30 Hz to 60 kHz |
(0.03 to 120) MHz |
3.4 % |
|
|
|
|
||||
|
|
|
|
|
|
|
|
(120 to 550) MHz |
3.4 % |
|
|
|
|
||||
|
|
|
|
|
|
|
|
(0.55 to 1) GHz |
3.2 % |
|
|
|
|
||||
|
|
|
|
|
|
|
(60 to 100) kHz |
(4.25 to 120) MHz |
3.4 % |
|
|
|
|
||||
|
|
|
|
|
|
|
|
(120 to 550) MHz |
6.8 % |
|
|
|
|
||||
|
|
|
|
|
|
|
|
(0.55 to 1) GHz |
13 % |
|
|
|
|
||||
|
|
0.2 |
kHz |
to |
0.3 |
kHz |
30 Hz to 60 kHz |
(0.03 to 120) MHz |
2.4 % |
|
|
|
|
||||
|
|
|
|
|
|
|
|
(120 to 550) MHz |
2.4 % |
|
|
|
|
||||
|
|
|
|
|
|
|
|
(0.55 to 1) GHz |
3.2 % |
|
|
|
|
||||
|
|
|
|
|
|
|
(60 to 100) kHz |
(4.25 to 120) MHz |
2.4 % |
|
|
|
|
||||
|
|
|
|
|
|
|
|
(120 to 550) MHz |
4 % |
|
|
|
|
||||
|
|
|
|
|
|
|
|
(0.55 to 1) GHz |
6.8 % |
|
|
|
|
||||
|
|
0.3 |
kHz |
to |
0.4 |
kHz |
30 Hz to 60 kHz |
(0.03 to 120) MHz |
2 % |
|
|
|
|
||||
|
|
|
|
|
|
|
|
(120 to 550) MHz |
2 % |
|
|
|
|
||||
|
|
|
|
|
|
|
|
(0.55 to 1) GHz |
4.9 % |
|
|
|
|
||||
|
|
|
|
|
|
|
(60 to 100) kHz |
(4.25 to 120) MHz |
2 % |
|
|
|
|
||||
|
|
|
|
|
|
|
|
(120 to 550) MHz |
3.1 % |
|
|
|
|
||||
|
|
|
|
|
|
|
|
(0.55 to 1) GHz |
4.9 % |
|
|
|
|
||||
|
|
0.4 |
kHz |
to |
0.5 |
kHz |
30 Hz to 60 kHz |
(0.03 to 120) MHz |
1.9 % |
|
|
|
|
||||
|
|
|